Section 3
3-23
If the condition :
(I(test) X (Rx/Rs)) Max Current
is satisfied, then the test is terminated.
Max current is the maximum current that can be carried by the REFERENCE
RESISTOR (Rs).
The reported ratio is (Rx/Rs).
3.7.2.2.1.1.2.<2 Terminal> MENU Fn-2
(Mode_Edit_Resistors_Setup_MAIN)
Enter configuration for 2-terminal high resistance mode measurements.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Resistance RS: ▒
▒Serial No. RS: ▒
▒Resistance RX: ▒
▒Reversal Rate: ▒
▒Test Voltage (V): ▒
▒Max Voltage (V): ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Edit │Ok │ │ ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Press <Edit> and use arrow keys to move the highlight to the data line to be changed.
Modify data and press <Ok>.
Where:
Test Voltage (V) is the common terminal voltage providing current for Rs
and Rx
and
Max Voltage (V) is the maximum voltage rating for Rs or Rx. Enter the
lower of the two maximum voltage ratings.