Section 3
3-24
3.7.2.2.1.1.3.<I Extend> MENU Fn-3
(Mode_Edit_Resistors_Setup_MAIN)
Enter range extender configuration data.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Resistance RS: 0000000000 ▒
▒Serial No. RS: 1111111111 ▒
▒Resistance RX: 2222222222 ▒
▒Reversal Rate: 3333333333 ▒
▒Extender Ratio: 444444444 ▒
▒ ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Edit │Ok │ │ ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Press <Edit> and use arrow keys to move the highlight to the data line to be changed.
Modify data and press <Ok>. Check resistance limits on this screen.
3.7.2.2.2.<Edit> Probes MENU Fn-2 (Probes_Setup_MAIN)
Display temperature probe configuration menu page 1 of 2.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Mode: RS : ▒
▒Display: <Fn-2> RS SN: ▒
▒Update: <Fn-3> R0: ▒
▒Filter: Pr SN: ▒
▒Power: RRate: ▒
▒ mAmps: ▒
▒══════════┬──────────┬─────────┬──────────▒
▒Mode │Display │Update │MORE 1-2 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Fahrenheit Update 1
kelvin Update 2
Ratio Update 4
Ohms
Celsius