Section 3
3-44
Function key <Fn-2>, <milliamps> sets the magnitude of the current output.
For the current to flow to the front panel <Mode> must be set to <4 Wire>.
The current range is from 0 to 150 mA.
Function key <Fn-3>, <Volts> sets the magnitude of the voltage output. For
the voltage to be set onto the front panel <Mode> must be set to <2 Wire>.
The voltage range is from 15.7 V to 990 V in steps of 15 V.
CAUTION
When changing the voltage level using this function the voltage must
first be set to zero for 4 seconds before setting the voltage to the other
value
Diagnostic menu selection page 3 of 4.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Set Internal Values
Reading: Range: <Fn-1>
Turns: Mode: <Fn-2>
Sign: Speed: <Fn-3>
milliAmps: Cur Ramp:
Volts: Volt Ramp:
▒──────────┬──────────┬─────────┬──────────▒
Range Mode Speed MORE 3-4
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Function key <Fn-1>, <Range> selects the measurement voltage range for
the built-in voltmeter from: <20V>, <2V>, <200mV>, <20mV>, <2mV>.
Function key <Fn-2>, <Mode> sets the measurement mode from the choice
of <2-terminal>, <4-terminal> and <I extend> (range extender).
Function key <Fn-3>, <Speed> sets the servo-loop speed from the choice of
<Slowest>, <Slow>, <Moderate> and <Fast>. The servo loop is used for
balancing the flux in the core.