Section 3
3-46
ERROR STATUS displays the error bits in hexadecimal notation.
╔═══════════════════════════════════════════════════════════════╗
0000_0000_0000_0000_0000_0000_0000
XXX XXXX XXXX XXX XXX ││││ X│││
││││ ││└ ROM 1 CHECKSUM
││││ │└─ ROM 2 CHECKSUM
││││ └── EEROM CHECKSUM
│││└───── +40 V RAIL
││└────── -40 V RAIL
│└─────── CALIBRATION
└──────── DAC A
└────────── DAC B
└─────────────── NANOVOLT DETECTOR
└───────────────────────────────── SELF TEST NOT FINISHED
╚═══════════════════════════════════════════════════════════════╝
NOTE: 0 = Success, 1 = Failure
ROM CHECKSUM is the sum of the crcs of both proms. EEROM CHECKSUM is the
CHECKSUM for the battery backed RAM.
3.7.4.6. <Chart> MENU Fn-3 (Sofcal_MAIN)
This menu selection allows display and edit of the chart recorder settings.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Chart Settings
PPM/V:
Ohms/V:
Deg/V:
Volts/V:
Cal Step(V):
▒──────────┬──────────┬─────────┬──────────▒
Select Ok
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
The chart recorder output consists of a 12 bit digital to analog converter having
4096 steps for a 10 volt dynamic range. The settings entered at this menu level
determine the measurement resolution of the chart recorder output. The parameter
<Cal Step (V):> sets the step calibration value that is issued to the chart recorder
output when the command <Cal Step> is made active during the measurement cycle
( with <Chart Cal> active). The value entered for <Cal Step (V):> will apply for the
active engineering units of measurement from the data set <ppm>, <ratio>,
<Celsius>, <Fahrenheit>, <kelvin>, <Ohms> and <Volts>.