  • J
    James SimpsonAug 17, 2025
    What to do if there is no sound from Gutermann Measuring Instruments?
    • J
      Jennifer JacksonAug 17, 2025
      If you are experiencing no sound, there might be several reasons. First, ensure that your BT headphones are properly paired. If they are, check the headset volume control and increase it if it's too low. Finally, make sure your BT headset is adequately charged; charge it regularly, similar to how you would charge a mobile phone.
  • D
    David DixonAug 20, 2025
    Why is there no radio connection to sensors on Gutermann Measuring Instruments?
    • A
      Ashley RossAug 20, 2025
      If there is no radio connection to the sensors, it could be due to low battery on the sensors. Charge the sensors using the car charger. Alternatively, the sensors might be turned off, so ensure that you switch them on.
  • S
    sanchezkellyAug 22, 2025
    What to do if Gutermann AQUASCAN 610 correlation doesn’t stop processing?
    • D
      Deborah JeffersonAug 22, 2025
      If the correlation doesn’t stop processing, it's possible that the correlation settings are on infinite. You should change the correlation settings or manually stop the correlation.
  • M
    michael54Aug 25, 2025
    Why Gutermann AQUASCAN 610 Measuring Instruments result is 0 or negative?
    • J
      Joel MoranAug 26, 2025
      A result of 0 or negative may occur if the leak noise originates from a leaking hydrant on which the sensor is placed, or beyond the sensor. In this case, fix the hydrant or move the sensor to bring the leak within the bracket.
  • S
    Samantha HayesAug 28, 2025
    Why Gutermann AQUASCAN 610 Measuring Instruments leak result is out by several meters?
    • J
      John ShieldsAug 28, 2025
      If the leak result is off by several meters, the incorrect velocity may be to blame, due to the wrong pipe material or age of pipe being entered. Perform a velocity check to resolve this.
  • J
    Joshua FlowersAug 30, 2025
    How to fix poor correlation spectra on Gutermann AQUASCAN 610 Measuring Instruments?
    • F
      Frank EdwardsAug 30, 2025
      Poor correlation spectra can result from poor sensor or radio contact. Try repositioning the sensors, cleaning the magnet contact area, or fitting an antenna extension cable.
  • N
    Nicole MckaySep 2, 2025
    What to do if Gutermann AQUASCAN 610 leak result is on a Tee section?
    • C
      Christina LopezSep 2, 2025
      If the leak result is on a Tee section, the leak could be up the branch of the Tee section. Try moving the sensor to the next section of the Tee.
  • L
    Lauren BaxterSep 5, 2025
    How to resolve Centre Correlation issues with Gutermann AQUASCAN 610 Measuring Instruments?
    • V
      vfergusonSep 5, 2025
      If you are experiencing centre correlation issues, it might be due to the sensors being on different pipes. Move the sensors to the same pipe. This issue may also occur with parallel mains, so ensure the sensors are on the same pipe in that situation as well.
  • R
    Ruth HaydenSep 8, 2025
    What causes corrupt display on Gutermann AQUASCAN 610 Measuring Instruments and how to fix it?
    • T
      Thomas Howard DDSSep 8, 2025
      A corrupt display is caused by memory corruption. To fix it, clear the memory.
  • J
    Jenna RaySep 12, 2025
    What to do if no display of correlator on Gutermann AQUASCAN 610 Measuring Instruments?
    • J
      Joe MeyersSep 12, 2025
      If there is no display of the correlator, the batteries are likely dead. Charge the unit.