EasyManua.ls Logo

GW Instek APS-7050 - Measure[:Scalar]:Power[:Ac]:Apparent; Measure[:Scalar]:Power[:Ac]:Pfactor; Measure[:Scalar]:Power[:Ac]:Reactive; Measure[:Scalar]:Power[:Ac][:Real]

GW Instek APS-7050
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
63
:MEASure[:SCALar]:POWer[:AC]:APParent
Query
Description
Returns the apparent power (VA).
Syntax
:MEASure[:SCALar]:POWer[:AC]:APParent?
Return
<NR2>
Returns the apparent power in VA.
:MEASure[:SCALar]:POWer[:AC]:PFACtor
Query
Description
Returns the power factor.
Syntax
:MEASure[:SCALar]:POWer[:AC]:PFACtor?
Return
<NR2>
Returns the power factor.
:MEASure[:SCALar]:POWer[:AC]:REACtive
Query
Description
Returns the reactive power (VAR).
Syntax
:MEASure[:SCALar]:POWer[:AC]:REACtive?
Return
<NR2>
Returns the reactive power in VAR.
:MEASure[:SCALar]:POWer[:AC][:REAL]
Query
Description
Returns the active power in Watts.
Syntax
:MEASure[:SCALar]:POWer[:AC][:REAL]?
Return
<NR2>
Returns the power in W.
:MEASure[:SCALar]:VOLTage[:RMS]
Query
Description
Returns the voltage (Vrms).
Syntax
:MEASure[:SCALar]:VOLTage[:RMS]?
Return
<NR2>
Returns the voltage in Vrms.

Table of Contents

Other manuals for GW Instek APS-7050

Related product manuals