OPERATION
119
Sets the NoGo condition to when
the input signal stays within the
limit boundary.
Sets the NoGo condition to when
the input signal exceeds the limit
boundary.
3. Press Violating to set the NoGo Action. This
setting determines the action of the scope when
the NoGo When setting is true.
The waveform stops when the
conditions are violated.
The waveform stops and a beep
will be output when the
conditions are violated.
Ignore violations and continue
to monitor the signal.
Output a beep when a violation
occurs, but continue to monitor
the signal.
4. Press Go-NoGo and turn Go-NoGo on.
An upper and lower template will appear on
the display as Ref1 and Ref2, respectively.
The Template Edit option will now be available
in the Go-NoGo menu.
5. The Template Edit option configures the upper
and lower template based on the source
channel selected above.
6. Select either the Max/Min or the Auto option