EasyManua.ls Logo

GW Instek GPT-700A Series - Page 7

GW Instek GPT-700A Series
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ELECTRICAL SAFETY TESTER
USER MANUAL
5
3) Insulation Resistance Specifications (for GPI-725A/ 735A/ 745A
only)
DC Voltage
50V/100V/500V/1000V
Resistance Accuracy
50V/100V:
1~50MΩ: ±5% of reading ±1 count
51~200MΩ: ±10% of reading ±1 count
201~2000MΩ: ±20% of reading±1 count
500V/1000V:
1~500MΩ: ±5% of reading ±1 count
501~2000MΩ: ±10% of reading±1 count
2001~9900MΩ: ±20% of reading±1 count
4) Ground Bond Specifications (for GPI-745A only)
Test Voltage Max. 8V(DC)
Current Range 3~42A
Current Resolution 0.02A
Current Accuracy
±1% of reading ±0.2A (3~8A)
±1% of reading ±50mA (9~42A)
Resistance Range
0~600mΩ
Resistance Resolution
0.1mΩ
Resistance Accuracy
±1% of reading ±3mΩ
Test Method Four Terminals Test Mode
5) Continuity Check Specifications (for GPT-705A/715A/ GPI-725A/
735A only)
Output Current Range 0.1A(DC)
Compliance Voltage Range 10%
Resistance Resolution
0.001Ω
ELECTRICAL SAFETY TESTER
USER MANUAL
6
6) Ramp Time and Test Time
AC Hi-Pot Ramp/Test time 000.0~999.9s Ramp/000.5~999.9 test
DC Hi-Pot Ramp/Test time 000.0~999.9s Ramp/000.5~999.9 test
Insulation Resistance Test
time
001.0~999.9s
Ground Bond Test Time 000.5~999.9s
Continuity Check Test Time 000.5~999.9s
7) ARC Detect
AC Detect Current 40 level (1~40mA)
DC Detect Current 10 level (1~10mA)
8) Storage
Groups
10
Steps
16
9) Interface
RS-232 Standard
GPIB Option
10) PLC Control
D-sub 9 pins female Standard
11) Scanner Interface
D-sub 9 pins female Standard
12) Real Plate Output
Hi-pot terminals and GB
terminals
Standard interface for GPI-745A only

Related product manuals