Hach AS950 - Quick Start Guide
Time Based Sampling Program
Initial Setup of New Sampler
1. Apply power to the sampler and turn on the controller.
2. Press MENU key on keypad
3. Navigate to General Settings
4. Select Date and Time
5. Enter the current date and time and select OK
6. Press MENU key on keypad
7. Navigate to Hardware Setup
8. Select Sampler
9. Select Rinses enter “1” and select OK
10. Select Sample Retries enter “1” and select OK
11. Press MENU key on keypad
12. Navigate to Programming
13. Select Programming
14. Navigate to and select Sampler Programming
15. Select Total Bottles enter “1” and select OK
16. Select Bottle Volume enter “3 gallon” and select OK
17. Select Tubing enterlength of intake tubing to the nearest foot and select OK
18. Select Pacing then select Time Weighted, select NEXT
19. With Time Weighted highlighted select Edit, enter “10 minutes select OK
20. Select Take First. Select Edit, select Immediately select Back
21. Select Sample Volume, select Fixed
22. Select Volume enter “50 ml”, select OK, select Back
23. Select Program Start, select Immediately on ‘Run’, select Next
24. Select Program End, select Edit, select After Samples, press Select
25. Highlight Samples, select Edit, enter “144” select OK, select Back
26. Select RUN/HALT key on keypad
27. Select Start Program
28. If warned about clearing data, affirm by selecting OK
29. Sampler is now Running
NOTE: Bold font in step 17 indicates a variable parameter that is site dependent. Each time the
sampler is moved to a new location you will want to enter the length of tubing between the intake
strainer and the sampler to the nearest foot. Entering the tubing length ensures that the sample
volume will be within tolerance.
AS950
Controller

Other manuals for Hach AS950

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach AS950 and is the answer not in the manual?

Hach AS950 Specifications

General IconGeneral
Sample FrequencyProgrammable
TypePortable
Sampling MethodVacuum Pump
Temperature ControlRefrigerated
Bottle Configuration24 x 1L bottles
Power Supply12V DC
CommunicationRS-232
Sampler TypePortable or Stationary
Vertical Lift8.2 m
Enclosure RatingIP67
Number of SamplesUp to 999
Storage Capacity24 samples
Bottle Options1L

Related product manuals