DOC346.53.80493
AS950 Controller
06/2016, Edition 8
Operations

Other manuals for Hach AS950

Questions and Answers

  • K
    Kevin BentonSep 12, 2025
    Why is the sample volume incorrect with my Hach Laboratory Equipment?
    • M
      Mario AllenSep 12, 2025
      The sample volume might be incorrect due to several reasons. Check the volume calibration and repeat if necessary. Ensure the correct tube length is specified in the sampler program. Make sure the intake tube is as vertical and short as possible. Install the shallow depth strainer if the current one isn't fully submerged. Replace worn pump tubing or the roller assembly. Verify the liquid detector is enabled and calibrate it with the sampled liquid.
  • B
    Benjamin ChenAug 22, 2025
    What to do if some measurements are not showing on the Hach Controller?
    • S
      Sabrina LambertAug 23, 2025
      Only the measurements recorded to the data log show. To see all the measurements supplied to the sampler by a sensor(s) connected to a sensor port, select Diagnostics>Sensor Ports. To show all the measurements supplied to the sampler by an external instrument(s) that is connected to the AUX I/O port, select Diagnostics>AUX and I/O Port.
  • I
    Isaac HernandezAug 27, 2025
    How to troubleshoot no instrument power in Hach Laboratory Equipment?
    • B
      Bradley HallAug 27, 2025
      If your Hach Laboratory Equipment has no instrument power, check the following: * Make sure that AC power gets to the electrical outlet. * If you have a refrigerated sampler, the power supply may be defective, so replace the power supply. * If the above steps do not solve the issue, the controller may be defective, so contact technical support.
  • A
    Andrew WarnerAug 28, 2025
    Why does Hach Laboratory Equipment sampler not have sufficient lift?
    • B
      Bradley SimmonsAug 28, 2025
      If the sampler does not have sufficient lift in your Hach Laboratory Equipment, consider the following: * The strainer may not be completely submerged, so install the shallow depth strainer (2071 or 4652). * The intake tube may have a leak, so replace the intake tube. * The pump tube may be worn, so replace the pump tubing. * The pump roller assembly may be worn, so contact technical support.
  • H
    Heather MichaelAug 29, 2025
    What to do if my Hach Laboratory Equipment has no instrument power?
    • J
      Jennifer ThomasAug 29, 2025
      If your Hach Laboratory Equipment has no instrument power, first ensure that AC power is reaching the electrical outlet. If the problem persists, it may be due to a defective controller.
  • J
    joshuabeanAug 31, 2025
    Why is the sample volume incorrect in my Hach Laboratory Equipment?
    • A
      Anne PhillipsAug 31, 2025
      If the sample volume is incorrect, several factors could be at play: * The volume calibration may be off, so repeat the volume calibration. * The tube length specified in the sampling program may be incorrect, so ensure it is correct. * The intake tube may not be purging completely; ensure it is as vertical and short as possible. * The strainer may not be fully submerged; install the shallow depth strainer (2071 or 4652). * The pump tubing or roller assembly may be worn; replace them as needed. * The liquid detector may be disabled; turn it on and complete a volume calibration. * The liquid detector may not be operating correctly; calibrate it with the sampled liquid.
  • J
    Jennifer HernandezNov 23, 2025
    What to do if the sample volume is not correct in Hach Laboratory Equipment?
    • S
      sara10Nov 23, 2025
      If the sample volume is not correct for your Hach Laboratory Equipment, there might be several reasons: * The volume calibration may be incorrect, so repeat the volume calibration. * The tube length specified in the sampling program may be incorrect, so make sure that the correct tube length is in the sampling program. * The intake tube may not purge completely, so ensure that the intake tube is as vertical and as short as possible. * The strainer may not be fully submerged, so install the shallow depth strainer (2071 or 4652). * The pump tubing and/or roller assembly may be worn, so replace the pump tube and/or roller assembly. * The liquid detector may be disabled, so turn the liquid detector on and complete a volume calibration. * The liquid detector may not operate prope...
  • B
    bwilsonSep 13, 2025
    How to troubleshoot no instrument power with AC on Hach Laboratory Equipment?
    • H
      Heather CummingsSep 13, 2025
      If there is no instrument power with AC, examine the AC power at the electrical outlet to ensure it is functioning correctly.
  • T
    Theresa HerreraSep 10, 2025
    What to do if the red indicator light flashes on the Hach Controller?
    • T
      Theresa LarsonSep 10, 2025
      If the red indicator light is flashing on your Hach Controller, it means that one or more measurements the sampler is configured to use (like pH and flow) are not available because the sensor(s) is not connected. To resolve this, connect the missing sensor to the sampler or ensure the sampler configuration is correct. Specifically, make sure that the sampler is not configured to record sensor measurements to the data log that are not available, or with alarms for sensor measurements that are not available.
  • C
    Catherine GarciaSep 6, 2025
    Why does no measurement data show on my Hach Controller?
    • M
      Megan HendricksSep 7, 2025
      If no measurement data is showing, it's likely because the data log is empty. Remember that measurement data is recorded to the data log according to the selected logging interval.

Summary

Product overview

User interface and navigation

Covers keypad, display, and main menu overview.

Keypad and display description

Explains the controller's keypad buttons and display elements.

Operation

Configuration

Sets general, hardware, sensor, data logging, and alarm configurations.

Sampling Program Setup

Configures sample collection parameters like type, pacing, volume, and distribution.

Calibration

Guides through sample volume, sensor, and temperature calibration.

Data Viewing and Management

Shows status, history, logs, system info, and USB operations.

Manual Operation

Controls pump, distributor arm, and grab samples for direct interaction.

Troubleshooting and Diagnostics

Addresses common issues and performs diagnostic tests.

Hach AS950 Specifications

General IconGeneral
Sample FrequencyProgrammable
TypePortable
Sampling MethodVacuum Pump
Temperature ControlRefrigerated
Bottle Configuration24 x 1L bottles
Power Supply12V DC
CommunicationRS-232
Sampler TypePortable or Stationary
Vertical Lift8.2 m
Enclosure RatingIP67
Number of SamplesUp to 999
Storage Capacity24 samples
Bottle Options1L