Why is the sample volume incorrect with my Hach Laboratory Equipment?
- MMario AllenSep 12, 2025
The sample volume might be incorrect due to several reasons. Check the volume calibration and repeat if necessary. Ensure the correct tube length is specified in the sampler program. Make sure the intake tube is as vertical and short as possible. Install the shallow depth strainer if the current one isn't fully submerged. Replace worn pump tubing or the roller assembly. Verify the liquid detector is enabled and calibrate it with the sampled liquid.