Spec Filter accuracy testing—DR 5000 and DR 2800
Test record template
Test Filter Wavelength Nominal value
1
1
Transfer the nominal values given in the quality control certificate HDQ178.
Tolerance Actual value
Stray light KV450/3 340 nm Abs
Photometric
accuracy
NG9/1 546 nm Abs ± 3%
NG5/2 546 nm Abs ± 3%
NG11/2 546 nm Abs ± 3%
Wavelength
accuracy
Ho 359–363 nm nm
± 1.5 nm
(DR5000)
± 2 nm
(DR2800)
BG20/2 805–809 nm nm