English 20
Please insert the USB
memory stick.
Insert a USB memory
stick into a USB A port on
the instrument.
Please check the
connection and contact
the administrator.
Network setup or FTP
error
Blank value correction
not possible!
Blank value correction
not possible with
LCW919.
File for instrument
update missing.
Error during update.
Check the USB memory
stick.
File for instrument
update is faulty.
Error during update.
Save the update file
again and repeat the
procedure.
It’s recommended to
execute a Full System
Check
Check of the air values
failed
Switch the instrument of
and then back on again.
If the system check is not
successful, contact the
manufacturer or his
representative.
Entry invalid! Password incorrect
Forgotten your
password?
Contact the
manufacturer or his
representative.
Absorbance > 3.5!
The measured
absorbance exceeds 3.5
Dilute sample and
measure again
Fault
Barcode check
number?
Please update program
data!
Deviation to the stored
data
Data updating
Error when calling up
the local IP address.
Network setup: DHCP
client has no connection
to the DHCP server
Enter the IP address
again.
Error displayed Definition Resolution
Error during default
gateway setup.
Network setup: default
gateway cannot be set
for fixed IP address
Try to create the
connection again.
Error during network
drive setup!
Error during network
setup
Check the settings.
Error during subnet
mask setup.
Network setup: Subnet
mask cannot be set for
fixed IP address
Enter the subnet mask
again.
Error copying from USB
memory stick.
Error during update
Start the procedure again
or contact the
manufacturer or his
representative.
Error in FTP
connection.
FTP error
Make sure that the
instrument is connected
to the network.
Fault
Program not
accessible.
Please update program
data!
Barcode test not present Data updating
Fault
Clean cuvette!
The cuvette is soiled or
there are undissolved
particles in the cuvette
Clean the cuvette; allow
the particles to settle
Fault
Test program stopped!
Please check lamp
Close the lid.
Error [xx]
Test program stops when
the instrument is started
Check the lamp and
replace it if necessary.
Close lid.
Press Start Again.
Fault
Test program stopped!
Please remove the
cuvette
Close the lid.
Test program stops when
the instrument is started
Remove the cuvette/
sample cell from the cell
compartment.
Press OK.
Error displayed Definition Resolution