42-19
MOLYBDENUM, MOLYBDATE, HR, continued
10. Place the blank
in the cell holder.
Note: Typical indoor
lighting permits the
DR/700 to operate with
the cell compartment
cover open. In bright
sunlight, it may be
necessary to close the
cell compartment cover.
Transfer 10 mL of the
blank solution to a
10-mL cell. If the 10-mL
cell is used for the blank,
another 10-mL cell must
be used for the sample.
11. Press: ZERO
The display will
count down to 0.
Then the display will
show 0.00 mg/L and
the zero prompt will
turn off.
13. Press: READ
The display will
count down to 0.
Then the display will
show the results in
mg/L molybdenum
(Mo
6+
).
12. Place the
prepared sample in
the cell holder.
Note: Typical indoor
lighting permits the
DR/700 to operate with
the cell compartment
cover open. In bright
sunlight, it may be
necessary to close the
cell compartment cover.
Transfer 10 mL of the
blank solution to a
10-mL cell. If the 10-mL
cell is used for the blank,
another 10-mL cell must
be used for the sample.