EasyManua.ls Logo

Hach DR/700 - NITROGEN, AMMONIA; Salicylate Method*

Hach DR/700
762 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61-43
Method 8155
NITROGEN, AMMONIA (0 to 1.00 mg/L NH
3
-N)
For water, wastewater, seawater
Salicylate Method*
1. Install module
number
61.01
in a DR/700.
Note: If samples cannot
be analyzed
immediately, see
Sampling and Storage
following these steps.
Adjust the pH of stored
samples before analysis.
2. Press: I/O
The display will show
610 nm
and module number
61.01
3. After 2 seconds,
the display will show
a program number,
the concentration
units and the zero
prompt. If necessary,
press UP arrow until
the lower display
shows program
number
61.06.1
4. Fill a 25-mL cell
to the 25-mL line
with sample.
Note: For proof of
accuracy, use a
0.20 mg/L NH
3
-N
solution (preparation
given in Accuracy
Check) in place of the
sample.
5. Fill a 25-mL cell
to the 25-mL line
with demineralized
water (the blank).
6. Add the contents
of one Salicylate
Reagent Powder
Pillow to each cell.
Cap and invert the
cells several times to
mix.
*Adapted from Clin. Chim. Acta. 1966, 14, 403