EasyManua.ls Logo

Hach DR/700 - Page 602

Hach DR/700
762 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61-46
NITROGEN, AMMONIA, continued
c) Analyze each sample as described above. The ammonia nitrogen
concentration should increase 0.08 mg/L for each 0.2 mL of standard
added.
d) If these increases do not occur, see Standard Additions (Section I) for
more information.
Standard Solution Method
Prepare a 0.20 mg/L ammonia nitrogen standard by diluting 2.00 mL of
the Nitrogen Ammonia Standard Solution (as NH
3
-N), 10 mg/L, to 100
mL with demineralized water. Or, using the TenSette Pipet, prepare a
0.20 mg/L ammonia nitrogen standard by diluting 0.4 mL of a Ammonia
Nitrogen Voluette Standard Solution, 50 mg/L as NH
3
-N, to 100 mL
with demineralized water.
STATISTICAL EVALUATION
A single operator repetitively tested samples of two laboratory prepared
solutions, using one DR/700, matched sample cells and two representative
lots of testing reagents, Testing 0.400 mg/L NH
3
as N concentration
samples, the standard deviation was ±0.006 mg/L NH
3
as N.
Testing zero concentration samples, the limit of detection was
0.029 mg/L NH
3
as N. The limit of detection was calculated as three
times the standard deviation when testing zero concentration samples
(Adapted from Analytical Chemistry,1980, 52, 2242-2249).
INTERFERENCES
The following ions may interfere when present in concentrations
exceeding those listed below:
Calcium 1000 mg/L as CaCO
3
Magnesium 6000 mg/L as CaCO
3
Nitrite 12 mg/L as NO
2
-
-N
Nitrate 100 mg/L as NO
3
-
-N
Orthophosphate 100 mg/L as PO
4
3-
-P
Sulfate 300 mg/L as SO
4
2-
Sulfide will intensify the color. Eliminate sulfide interference as
follows:
a) Measure about 350 mL of sample in a 500-mL erlenmeyer flask.