32
Display, Alarms and Self-Diagnonis
Connect the power  supply for the refrigerator and switch the  power switch to the “ON” position to 
enter into power-on state.
1.Working status prompt of the display panel
•“Upper” indicator light:  when the light is ON, the digital tube  displays the temperature of the 
upper sensor.
•“Lower” indicator light:  when the light is ON, the digital tube  displays the temperature of the 
lower sensor. (When the upper and lower temperature sensor indicators are ON at  the same 
time, it displays the average value of the upper and lower temperature sensors).
•“Power” indicator light: when the light is ON, it indicates that it is in the main power supply state; 
when the light ickers, it indicates that it is in the main power failure state.
•“Alarm” indicator  light:  when the  light  flickers,  it  indicates that  there  is a  fault or  abnormal 
condition.
•“Door open” indicator light: when the door is opened, this light is ON; when the door alarm is 
activated, the light ickers; when the door is closed, the light is OFF.
•“Hot condenser” indicator light: when the light ickers, it indicates that the condenser is too dirty 
and needs to be cleaned in time.
•“Network” indicator light: when the light is ON,  it  indicates  that  it  is in the networking state; 
during communicating, the light ickers.
•“NFC” indicator light:  when  the light is  ON,  it indicates that  the  NFC card reading  board  is 
connected; when the light is OFF, it indicates that the NFC recording board is not connected; 
when the light ickers, it indicates that the door lock is opened.
2.Alarms and Self-Diagnonis 
Alarm Code Status Indication Buzzer
Remote 
alarm
High 
temperature 
alarm
E00
Temperature of the upper / 
lower temperature sensor is 
≥ the set temperature of high 
temperature alarm for one 
minute
Alarm light ickers
Makes 
intermittent 
alarm sound
Start up
Low 
temperature 
alarm
E01
Temperature of the upper / 
lower temperature sensor is 
≤ the set temperature of low 
temperature alarm for one 
minute
Alarm light ickers
Makes 
intermittent 
alarm sound
Start up
Power failure 
alarm
E30
Power failure of the 
refrigerator
The alarm light and 
power indicator light 
icker, and the digital tube 
intermittently display the 
temperature.
Makes 
intermittent 
alarm sound
Start up
Power board 
fault alarm
E31 12V voltage < 9V is detected Alarm light ickers
Makes 
intermittent 
alarm sound
Start up
LED Type Display (HYC-509/509F/509R/1099/1099F/1099R)