halstrup-walcher PU10 - Electrical Data; Physical Data; Appendix A: Materials

Default Icon
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PU/PI and PS10 Instruction Manual
11
Electrical data
power consumption
max. 1.2 W
supply voltage
24 VDC +20% / -15%
(filtered, permissible peak-to-valley ratio = 1000 mV)
230VAC, 115VAC, 24VAC +6 % / -15 %, 50/60 Hz
(optional)
minimum load resistance R
L
R
L
2 k for an output voltage of 0...10 V
R
L
5 k for an output voltage of 5 V
R
L
10 k for an output voltage of 10 V
for models PU and PS10 with output voltage
maximum effect of load resistance = 0.3%
maximum output load R
B
R
B
= 500
for models PI and PS 10 with current output
output load dependence: < 0.3 %
display
3½ or 4½-place LCD, character height = 13 mm
(optional)
output signal
0 to 10 V, 0 to 20 mA or 4 to 20 mA; ±5 V and ±10 V are
also possible
Physical data
pressure port
Ø 6.5 mm for NW5 tubing
(interior tubing diameter = 5 mm)
electrical connections
screw terminals for cables up to 2.5 mm
2
mounting orientation
vertical
(when placing your order, please indicate if a horizontal
orientation is required)
dimensions (w x h x d)
80 x 120 x 73 mm
120 x 122 x 75 mm (for LCD option)
protection class
IP54
weight
between 300 g and 700 g, depending on design
options
-place LCD
-place LCD
linearity protocol
DKD calibration certificate
cable gland: PG9 or PG11 threaded connections
output signal attenuation up to 5 s
silicon-free materials for parts in contact with medium
Appendix A: Parts in contact with measurement medium
Beryllium bronze CuBe2 Araldite CY236 / HY988
Mu metal (nickel alloy) Loctite 242e
Brass CuZn39Pb3 Carbonyl iron
Aluminum AlCuMgPb / AlMg3 KEL (FPM: fluorinated rubber)
Silicon (tubing) optional: Viton Vepuran Vu 4457/51