BSM Series
Electronic Precision Balances

Questions and Answers

  • A
    Alex AvilaAug 3, 2025
    Why is the weight readout wrong on my Hanchen Scales?
    • M
      Mary RiveraAug 3, 2025
      If the weight readout on your Hanchen Scales is obviously wrong, it could be due to a couple of reasons. First, the balance might not have been calibrated or adjusted before weighing, in which case you should calibrate or adjust the balance. Second, the balance might not have been zeroed before weighing; make sure to zero the balance before you start weighing.
  • J
    Jonathan WilkinsonAug 7, 2025
    What to do if no segments appear on the display of my Hanchen BSM?
    • E
      Eric GarrettAug 8, 2025
      If no segments appear on the display of your Hanchen Scales, there are a few potential causes. First, check the AC power supply and ensure the power supply is plugged in. Second, the load might exceed the balance capacity, so try unloading the balance. Finally, ensure nothing is touching the weighing pan; if something is, move it.
  • C
    Caitlin LeAug 13, 2025
    What to do if my Hanchen BSM display E22?
    • L
      lindarobinsonAug 13, 2025
      If your Hanchen Scales are showing an E22 error, it means the poise weight is not enough. To resolve this, load on enough poise for weighing.
  • T
    thomas39Aug 18, 2025
    What does E11 mean on Hanchen Scales?
    • A
      Angela GutierrezAug 18, 2025
      If your Hanchen Scales are showing an E11 error, it indicates that the calibrate weight is wrong. To fix this, load weight on the poise to recalibrate.
  • M
    Mark Bowman MDAug 22, 2025
    What does E12 mean on Hanchen Scales?
    • S
      Sheila GibsonAug 22, 2025
      If your Hanchen Scales are showing an E12 error, it means the calibrate weight is unstable. You should eliminate the unstable factor and then recalibrate.
  • E
    ekaiserAug 28, 2025
    How to fix if maximum capacity is less than the data list in the Technical specification of Hanchen BSM?
    • L
      Laura BakerAug 28, 2025
      To address this issue, switch on the Hanchen balance without the weighing pan. Then, place the weighing pan and restart the balance while pressing the ?? button.

Summary

Introduction and Safety

Application Overview

Introduces the BSM series balances and their general capabilities.

Safety Precautions

Critical instructions to ensure safe operation and prevent equipment damage.

Getting Started

Guidance on unpacking, installation, conditioning, and leveling the balance.

Balance Operation

Display and Control Elements

Identifies and explains the balance's display indicators and control buttons.

Basic Weighing

Step-by-step instructions for performing simple weight measurements.

Calibration and Adjustment

Procedures for ensuring balance accuracy through calibration.

External Calibration

Detailed steps for calibrating the balance using external weights.

Application Programs

Counting Program

How to use the counting function for piece measurement.

Weighing in Percent Program

Procedure for obtaining weight readings as a percentage.

Density Determination Program

Method for calculating the density of solid objects.

Animal Weighing;Averaging

Technique for weighing unstable samples by averaging.

Toggling Weight Units

How to switch between different units of weight measurement.

Configuration and Support

Balance Settings and Parameters

Navigating menu codes and understanding parameter settings.

Troubleshooting Guide

Solutions for common error codes and operational issues.

Care and Maintenance

Procedures for cleaning, repairs, and safety checks.

Hanchen BSM Series Specifications

General IconGeneral
BrandHanchen
ModelBSM Series
CategoryScales
TypePrecision Balance
CalibrationExternal Calibration
Tare FunctionYes
Overload ProtectionYes
Readability0.001g
Repeatability±0.001g
Linearity±0.002g
Platform SizeVaries by model
MaterialStainless Steel
DisplayLCD
Unitsg, oz, lb, ct
InterfaceRS232