2-2. Location of placement
To be installed on the rigid and flat level;
The rotation of heavy rotor in the long period on the inclined level may cause
the deformation of rotating shaft.
Smooth air ventilation;
Centrifuge needs 30cm of vacant space of front & rear and right & left. An airway
should not be covered by the another instrument or fabric, avoid the place of
heavy dust.
Constant temp. & humidity;
The sensitive electronic controller is affected by ambient temp. and humidity.
Avoid direct sunray and heater and keep the constant temp. & humidity.
No erosive gas;
Avoid the place of erosive gas. Contacting sulfurous acid/ chlorine gas erodes
Rotor & shaft and damage metal parts.
Leveling;
Shaft should be vertical exactly and Centrifuge is to be placed horizontally
with flatness measuring device.