Operation Manual
Smart R17
Micro Centrifuge

Questions and Answers

  • A
    Andrew PriceAug 17, 2025
    What to do if Hanil Laboratory Equipment shows an imbalance error?
    • C
      Chelsea SmithAug 17, 2025
      An imbalance error can occur in Hanil Laboratory Equipment due to several reasons. It may be caused by dissymmetrical placement of tubes, irregular sample volumes, or a loosened rotor. To resolve this, check the tightness of the rotor loading and ensure proper tube loading. Also, verify the equipment is on a stable surface.
  • M
    melissa89Aug 19, 2025
    Why does my Hanil Laboratory Equipment show an over speed error?
    • S
      samanthajimenezSep 12, 2025
      An over speed error in Hanil Laboratory Equipment can occur if the speed exceeds 2% of the set RPM or due to incorrect tuning of the motor and controller. Try switching the power off and on, then restart the equipment. If the problem persists, check the controller and motor.
  • B
    Brian BlairAug 22, 2025
    What to do if my Hanil Smart R17 Laboratory Equipment shows a system error?
    • C
      christina54Aug 22, 2025
      A system error on Hanil Laboratory Equipment can be caused by an error in the controlling system. In this case, switch the power off and then restart the equipment.
  • G
    Gloria JonesAug 25, 2025
    Why is my Hanil Smart R17 motor overheating?
    • J
      Jessica SanchezAug 26, 2025
      The motor of your Hanil Laboratory Equipment may overheat due to a temperature increase inside the motor. Check the stator of the motor. As an alternative, switch the power off, and restart the equipment after one hour.
  • A
    Alicia LynnAug 27, 2025
    What to do if Hanil Smart R17 Laboratory Equipment speed is low?
    • M
      Mckenzie BoydAug 28, 2025
      If your Hanil Laboratory Equipment is running at a low speed, not reaching 200 RPM within 2 seconds, check the controller and motor. You can also try switching the power off and then restarting the equipment.
  • L
    Luke GreenAug 30, 2025
    How to reset the over temperature limit on Hanil Laboratory Equipment?
    • J
      John KingAug 31, 2025
      If your Hanil Laboratory Equipment displays an 'Over Temp.' error, it means the temperature has exceeded the preset limit. Reset the preset value of the limit, which corresponds to the 'L-Cal' value in the preset menu.
  • C
    Candace RichardsonSep 2, 2025
    What to do if Hanil Smart R17 Laboratory Equipment shows an under speed error?
    • A
      Angela RamseySep 2, 2025
      If your Hanil Laboratory Equipment displays an under speed error, meaning the speed is 2% below the set RPM during operation, switch the power off and restart the equipment.
  • K
    krodgersSep 5, 2025
    What to do if Hanil Smart R17 won't start and shows 'Door Open'?
    • L
      Lauren WilsonSep 5, 2025
      If your Hanil Laboratory Equipment displays a 'Door Open' error upon starting, it indicates that the door is open during operation. Check the door-in switch.
  • A
    Amanda Wilson PhDSep 7, 2025
    What to do if the Hanil Laboratory Equipment door won't open after stopping?
    • B
      bondtimothySep 7, 2025
      If the door of your Hanil Laboratory Equipment does not open after the device has stopped, check the door-in switch.
  • W
    William MeyerSep 12, 2025
    What to do for 'Door Close' error on Hanil Laboratory Equipment?
    • K
      Kara ParksSep 12, 2025
      If your Hanil Laboratory Equipment shows a 'Door Close' error, indicating the door is not closed, check the door-in switch.

Hanil Smart R17 Specifications

General IconGeneral
TypeRefrigerated Centrifuge
DisplayLCD
Max. RPM17000 rpm
Temperature Range-20°C ~ 40°C
Safety DevicesImbalance sensor, Lid lock