Conductivity Meters with Automatic Temperature Compensation
ISTCONMET 05/05
Specifications Conmet
®
1 Conmet
®
2 Conmet
®
3
Range 0 to 1999 µS/cm 0.00 to 19.99 mS/cm 0.00 to 19.99 ppt (g/L)
Resolution 1 µS/cm 0.01 mS/cm 0.01 ppt (g/L)
Accuracy (@20°C/68°F) ±2% F.S.
Calibration Manual, 1 point, through slope trimmer
Temp. Compensation Automatic, 0 to 50°C, with ß=1.9% ±0.15% /°C, and 25°C (77°F) reference temperature
Probe (included) HI 3291 HI 3292 HI 3292
Battery Type / Life 4 x 1.5V alkaline / approx. 150 hours of continuous use
Environment 0 to 50°C (32 to 122°F); 95% RH max.
Dimensions / Weight 265 x 29 x 15 mm (10.4 x 1.1 x 0.6’’) / 400 g (0.88 lb.) complete with the carrying case
Each meter is supplied in a rugged plastic carrying case with 4-ring conductivity probe, calibration solution bottle (120 mL), screwdriver and instructions.
Operations: • Connect the conductivity probe to the meter (Fig. 1) and turn the unit on by sliding the ON/OFF switch on the top of the instrument.
• Dip the probe into the solution by ensuring that all 4 rings of the probe are immersed and without exceeding the maximum level (Fig. 2).
• Stir gently to remove any air bubbles trapped inside the probe and wait for the reading to stabilize.
• To minimize errors, it is recommended to rinse the probe with some of the sample to be tested before taking measurement.
• After use, switch the unit off, rinse the probe in alcohol for a few minutes.
• If the display fades, replace the batteries. Switch the unit off, pull out the battery compartment and replace the batteries (Fig. 3).
Battery replacement must only take place in a non hazardous area using the battery type specified in this instruction manual.
Calibration: • Dip the probe into the proper (supplied) calibration solution and wait for the reading to stabilize.
• If necessary, correct the reading to the solution value by slowly turning the calibration trimmer with the supplied screwdriver.
These meters are in
compliance with the
CE directives
Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments Conmet1 (HI 98305 and is the answer not in the manual?

Hanna Instruments Conmet1 (HI 98305 Specifications

General IconGeneral
ModelHI98305
Conductivity Range0 to 1999 µS/cm
Resolution1 µS/cm
Accuracy±2% F.S.
Temperature CompensationAutomatic, 0 to 50°C (32 to 122°F)
Operating Temperature0 to 50°C (32 to 122°F)
Dimensions175 x 41 x 23 mm (6.9 x 1.6 x 0.9")
Environment0 to 50°C (32 to 122°F)

Summary

Conductivity Meter Specifications and Operation

Product Specifications

Details technical specifications for Conmet 1, 2, and 3 meters including range, resolution, accuracy, calibration, compensation, probe, battery, environment, and dimensions.

Meter Operations Guide

Provides instructions on connecting the probe, immersing it in solution, stirring, rinsing, and switching off the unit after use.

Battery Replacement

Outlines the process for replacing batteries, including switching off the unit and accessing the compartment.

Calibration Procedure

Explains how to calibrate the meter by dipping the probe into calibration solution and adjusting the reading.