Types of Logging
Interval logging: A continuous log recorded using a user‑selected timed period. (This is not
available in Basic mode).
Manual log on demand: Readings are logged each time LOG key is pressed. All of the records
are stored in a single Manual lot for the measurement type. New records made on different days are
stored in the same Manual lot.
Manual Stability log on demand: A log on demand that is made each time LOG key is pressed
and the stability criteria is reached. Stability criteria may be set to Fast, Medium or Accurate settings.
In Setup mode, choose log parameter, press CAL/MODIFY key then use RANGE/u key to select
between Interval, Manual, or Stability. When Interval is displayed, use key to select the
setting for the timed interval. When Stability is displayed, use the key to select the
measurement stability setting.
LOGGING FUNCTION
Each time an Interval log is initiated, a new lot is created. The
maximum number of Interval lots that may be stored is 100. If a 101
st
lot is attempted, “MAX LOTS” will be displayed. Some lots will need to
be deleted. The lot numbering is up to 999 and restarts if all lot logs
are deleted. All log on demand and stability log on demand are stored
in a single lot. The maximum number of records that may be stored in
a Manual or Stability lot is 200 records.
SETUP/ INSTALLATION
If the log memory is full during any logging session, the “LOG FULL“
message will be displayed on the third LCD line for a brief moment
and logging will cease. The display will return to the measurement
screen.
Logging type is configured in SETUP.