7
6
FUNCTIONAL DESCRIPTION HI 8314FUNCTIONAL DESCRIPTION HI 8314
FUNCTIONAL DESCRIPTION HI 8314FUNCTIONAL DESCRIPTION HI 8314
FUNCTIONAL DESCRIPTION HI 8314
1) BNC socket for combination pH or ORP electrodes.
2) LCD display.
3) mV to display the mV (ORP) readings when using an ORP
electrode or the mV equivalent to pH values when using a
pH electrode.
4) STD trimmer for OFFSET calibration of pH.
5) Phono plug socket for temperature probe HI 7669AW.
6) ON/OFF to switch the instrument on or off.
7) pH to display the pH value.
8) °C to display the temperature measurement (when the
temperature probe is disconnected, the LCD will show 25°C).
9) SLOPE trimmer for SLOPE calibration of pH.
1
5
2
7
3
8
4
9
6
pH
ON
OFF
mV
°C
pH
HI 8314
membrane
pHmeter
STD
1nd
Calibration at pH
6.86 or 7.01
SLOPE
2nd
Calibration at pH
4.01 or 10.01
CALIBRATION
HI 8014
Range pH 0.00 to 14.00
mV 0 to ±1999
Resolution pH 0.01
mV 1
Accuracy pH ±0.01
(@20°C/68°F) mV ±1
Typical EMC pH ±0.05
Deviation mV ±8
pH Calibration Manual 2 point through
offset and slope trimmers
Offset Calibration ±1 pH
Slope Calibration from 85 to 105%
Temperature Manual from 0 to 100°C
Compensation (32 to 212°F)
Electrode HI 1332B combination, double-junction,
refillable pH electrode BNC
and 1 m (3.3') cable (included)
Input Impedance 10
12
ohm
Battery Type 1 x 9 volt
Life 100 hours of continuous use
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Dimensions 185 x 82 x 45 mm
(7.3 x 3.2 x 1.8")
Weight 520 g (1.1 lb.)
SPECIFICATIONS HI 8014SPECIFICATIONS HI 8014
SPECIFICATIONS HI 8014SPECIFICATIONS HI 8014
SPECIFICATIONS HI 8014