Hanna Instruments HI8417 User Manual

Hanna Instruments HI8417
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
21
Procedure:
Switch the instrument on after connecting the pH
electrode.
Rinse and immerse the pH electrode in pH 7.01
buffer and shake briefly. Wait 1 or 2 minutes for
the reading to stabilize.
Note the temperature of the buffer solution using
a ChecktempC or a glass thermometer (e.g.20°C).
Press the °C key to select temperature setting. Ad-
just the Temperature °C knob until display shows
the noted temperature.
Press the pH key to select pH measure-
ment.
Wait 1-2 minutes and adjust the OFFSET
knob until display shows the pH value at
the noted temperature (see the pH versus
temperature chart at page 23).
Rinse and immerse the pH electrode in
pH 4.01/pH 10.01 buffer and shake
briefly.
Wait 1-2 minutes and adjust the SLOPE
knob until display shows the pH value at
the noted temperature (see the pH versus
temperature chart at page 23).
The pH calibration is now complete.
PH CALIBRATION FOR HI8520 AND HI8521
Procedure:
Switch instrument on after connecting the pH elec-
trode and the temperature probe.
Immerse the pH electrode and the temperature
probe in pH 7.01 buffer solution, shake briefly
and wait 1 or 2 minutes for thermal equilibrium.
Press the RANGE key to display pH measurement.
pH
OFFSET
pH
SLOPE
°
C
°C
ON
OFF
°C
TEMPERATURE °C
°C
pH
ON
OFF
4 cm
(1½
")
RANGE