11 Setup
pH Alarm (Options / Default)
pH High Alarm
Enabled (En), Disabled (dIS)
(Enabled)
En
P H H I G H
dIS
P H H I G H
pH High Value
0.1 to 14.0 pH *
(8.6 pH)
8.6
P H H I G H
8.8
P H H I G H
With pH High Alarm configured as enabledPH HIGH VALUE” is scrolled on the bottom of the LCD.
Actual pH value (editable) is displayed blinking.
Press
e to enter edit mode and use a / v to change value. Press e to save.
pH Low Alarm
Enabled (En), Disabled (dIS)
(Enabled)
En
P H L O W
dIS
P H L O W
pH Low Value
0.0 to 13.9 pH *
(7.7 pH)
7.7
P H L O W V
7.9
P H L O W V
With Low Alarm configured as enabledPH LOW VALUE” is scrolled on the bottom of the LCD.
Actual pH value (editable) is displayed blinking.
Press
e to enter edit mode and use a / v to change value. Press e to save.
* Available range changes based upon other settings. High alarm value must be set higher than low alarm
(e.g. if low alarm was set 7 pH, then high alarm range is 7.1 to 14 pH).