Quick start 
    EN 
HDP10000 series user manual                                                                              Copyright Qingdao Hantek Electronics Co., LTD                   
12 
Press the    to enable the OCP function. The indicator light will be on. Press   
again to disable the OCP function, the indicator will be off. Long press    to enter 
the setting mode, use the current value adjustment knob to change the OCP set value. 
When the actual output current value is higher than the OCP set value, the protection 
function will be triggered and the screen will display  . The output is forcibly turned 
off, but the ON/OFF indicator light is not turned OFF. Press    to clear the error and 
turn off the indicator light. 
11. Voltage value adjustment knob 
It is the same as the current value adjustment knob. 
12.  
Press the    to enable OVP function. The indicator light will be on. Press   
again to disable OVP function. The indicator light will be off. Long press    to enter 
the setting mode and use the voltage adjustment knob to change the OVP set value. 
When the actual output voltage value is higher than the OVP set value, the protection 
function will be triggered and    will be displayed on the screen. The output is 
forcibly turned off, but the ON/OFF indicator light is not turned off. Press    to clear 
the errors and turn off the indicator light. 
13. CC constant current output state 
Under constant current working state, the status indicator will be on. 
14. Button lock   
Press  . The indicator light will be on. In button lock mode, only    and 
can be used. Press the    again to quit this mode. 
 
15. CV constant voltage output state 
Under constant voltage working state, the status indicator will be on.