2.0 WorkAreaRequirements
The Reagent Supply Kit (Cat. No. HP1SK) should be positioned at the same level as the
instrument on the counter immediately behind the instrument. The HemaPRO has a flat base in
the rear on which the reagents may be positioned.
NOTE: Do not place reagents above or below the instrument.
The HemaPRO uses an external power supply module that requires an input of 50-60
Hz, 100-240V, 0.8 A. This unit requires power from a grounded outlet. The HemaPRO should
only be used with the provided power cord. The power supply is a Class I supply that must be
connected to an earthed (grounded) main power outlet. Failure to connect the HemaPRO as
specified will prevent the electrical safety protection features to function as designed.
The HemaPRO has one waste tube which drains all of the waste fluids. The HemaPRO
drain line can be placed into a customer designated disposal container to accommodate the
waste fluids according to your facility and local regulations.
The HemaPRO is pre-installed with a tubing set to connect the instrument to each
reagent. Each reagent supply line is equipped with a cannula (rigid tube) to submerge in the
specified reagent supply container. Labels attached to those lines indicate which reagent is to
be used with each line.