Error Possible causes Remedy
Implausible measured
values
Implausible measured values often occur
due to inappropriate handling of the
device or mistakes during the measuring
procedure.
Please refer to Chapter 5 ‘Measuring
your blood pressure’ as well as the
safety information. Then repeat the
measurement.
Switch the device off if an error symbol appears. Check for all possible causes and follow the 10 golden rules
(Chapter 5 ´Measuring your blood pressure`) as well as the instructions for self-measurement from Chapter 2
‘Important information’.
Relax for one minute and then take the measurement again.
9. Monitor maintenance
Only use a soft, damp cloth to clean the device. Please do not use thinner, alcohol, detergents or solvents.
The cuff can be cleaned carefully using a slightly damp cloth and mild soap solution. Do not completely immerse the
cuff in water.
It is recommended to clean and disinfect the cuff regularly or after each use, especially when used by several users,
to prevent infection. The cuff should be disinfected, particularly on the inside, by wiping with a disinfectant. Use
a disinfectant that is compatible with the cuff materials, e.g. 75 % ethanol or isopropyl alcohol. To protect from
external influences, keep the device, the cuff and these instructions in the storage bag.
If you store a device and cuff, do not place any heavy objects on the device and the cuff. Remove the batteries.
10. Accessories
To ensure the accuracy of the measurement, please use only original accessories from HARTMANN, such as the
Veroval®
power supply unit (model number: 925 391), which can be ordered from your pharmacist or medical
specialist.
Mains power mode: On the back of the unit is a connector socket for the adapter for mains operation
(output 6V DC/600 mA). Only use the
Veroval®
power supply unit. The warranty covering the function and measuring
accuracy of the monitor cannot be assumed otherwise.
For more information, please visit www.veroval.info.
11. Warranty conditions
We offer a 3-year warranty for this high-quality blood pressure monitor from the purchase date and in accordance
with the conditions listed below.
Warranty claims must be made during the warranty period. The purchase date must be documented by an
appropriately completed and stamped warranty certificate or proof of purchase.
Within the warranty period, HARTMANN will replace or repair free of charge any faulty device components caused by
material faults or manufacturing errors. This does not extend the warranty period.
For local warranty conditions, please consult the local warranty certificate.
This device is solely intended for the purpose stated in these instructions for use.
The manufacturer is not liable for damage or injuries resulting from improper handling, use or unauthorized
interference. Accessory parts that are subject to wear and tear (batteries, cuffs, etc.) are excluded from the warranty.
Claims for compensation are limited to the value of the product; compensation for consequential damage or injury is
expressly excluded.
In warranty cases, please send the monitor with cuff and, if applicable, the power supply unit, together with the fully
completed and stamped warranty certificate or purchase receipt directly or via your dealer to the Customer Services
department of your country.
12. Contact information for customer enquiries
ZA – HARTMANN South Africa
Epsom Avenue, Northriding,
2169 Johannesburg
www.veroval.info
Tel. +27 860 4278 6266
phzahelpdesk@hartmann.info
If necessary, please contact us at the relevant address listed at the end of the instructions for any questions regarding
initial startup, use and maintenance of the device, or to report an unexpected operating malfunction or incident. For
a patient/user/third party in the European Union and in countries with identical regulatory regime (Regulation (EU)
2017/745 on Medical Devices); if, during the use of this device or as a result of its use, a serious incident has occurred,
please report it to the manufacturer and/or its authorized representative and to your national authority.
Vigilance contact point for Europe: www.ec.europa.eu/growth/sectors/medical-devices/contacts/
13. Technical data
Model Veroval® BPU 26
Type GCE609
Measuring method Oscillometric
Display range 0–300 mmHg
Measuring range Systolic (SYS): 50–280 mmHg
Diastolic (DIA): 30–200 mmHg
Pulse: 40–199 pulse beats per minute
Displaying of correct values outside the measuring range cannot be guaranteed.
Display unit 1 mmHg
Technical measurement
accuracy
Cuff pressure: ±3 mmHg
Pulse: ±5 % of the displayed pulse rate
Clinical measuring
accuracy
Complies with the requirements of IEC 80601-2-30;
Korotkoff validation method: Phase I (SYS), Phase V (DIA)
Operating mode Continuous mode
Nominal voltage DC 6V
Power supply 4 x 1.5 V alkaline manganese (AA/LR06) batteries or optional Veroval® power
supply unit
Expected service life 20,000 measurements
Battery capacity Approx. 800 measurements
Protection against electric
shock
Medical-electrical equipment with internal power supply (only when batteries are
used); applied part: type BF
Protection against harmful
ingress of water or solid
materials
IP21
Inflation pressure Comfort Air Technology:
Individual inflation pressure depends on systolic blood pressure
Automatic switch-off function 2 minutes after measurement/otherwise 30 sec.
Cuff Veroval® cuff for BPU 26, cuff for arm circumference of 22 to 42 cm
Model number: 30602
Memory capacity 2 × 100 measurements with average of all measurements and morning/evening
average of the last 7 days
Operating conditions Ambient temperature: +10 °C to +40 °C
Relative humidity: <90 %, non-condensing
Air pressure: 800 hPa – 1050 hPa
Storage/transport
conditions
Ambient temperature: –20 °C to +55 °C
Relative humidity: <90 %, non-condensing
Serial number On the rating label
Reference to standards IEC 60601-1; IEC 60601-1-2 (In accordance with CISPR11,
IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-8)
Weight Approx. 220 g (without batteries)
Dimensions 134(L) x 91(W) x 48(H) mm
Data transfer 2400 MHz - 2483.5 MHz frequency band
Transmission power: +5 dBm
The blood pressure monitor used Bluetooth® low energy technology compatible
with Bluetooth® ≥ 4.0 smartphones/tablets
Power supply unit:
Model no. LXCP12-006060BEH
Input 100–240V~, 5060Hz, 0.5A max
Output 6V DC, 600mA
Manufacturer Globalcare Medical Technology Co., Ltd.
Protection The device is double-insulated and has a primary fuse which disconnects the
monitor from the mains in the event of a fault. Make sure you have removed the
batteries from the battery compartment before connecting the power supply
unit.
Polarity of the DC voltage connection
Protective insulation / Protection Class 2
Housing and protective covers The power supply unit housing protects against touching parts that can carry a
current (fingers, needles, test hooks). The user must not touch the patient and the
outlet plug of the AC / DC power supply unit at the same time.
Legal requirements and guidelines
The Veroval® BPU 26 conforms with the Regulation (EU) 2017/745 of the European Parliament and of the Council on
medical devices, as well as the respective national provisions, and bears the CE mark.
The device complies with international standard IEC 80601-2-30 (Medical electrical equipment Part 2-30: Particular
requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers).
This product is subject to special precautionary measures with regard to electromagnetic compatibility.
Beyond statutory requirements, the device has been clinically validated by the ESH-IP2 Protocol of the European
Society of Hypertension (ESH) and the protocol of the German society of hypertension (DHL).
We hereby confirm that this product complies with the European RED Directive 2014/53/EU. The CE Declaration of
Conformity for this product can be found under: www.veroval.info
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth® SIG, Inc. and any use of such marks
by PAUL HARTMANN AG is under license. Other trademarks and trade names are those of their respective owners.
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service
mark of Apple Inc., registered in the U.S. and other countries.
Google Play and the Google Play logo are trademarks of Google LLC.
Android is a trademark of Google LLC.
Date of revision of the text: 2022-08-11, Version 1
Subject to errors and changes
Upper arm blood pressure monitor
Purchase date
Serial number (see rating label)
Reason for complain
Dealer‘s stamp
Globalcare Medical Technology Co., Ltd.
7th Building, 39 Middle Industrial Main Road
European Industrial Zone, Xiaolan Town
528415 Zhongshan City, Guangdong Province, P.R.C.
www.globalcare.com.hk
Beurer GmbH
Söflinger Str. 218
89077 Ulm, Germany
Camara and Partners Sàrl
Route de St Cergue 14
1260 Nyon, Switzerland
Donawa Lifescience
Piazza Albania, 10
00153 Rome, Italy
www.donawa.com
925106/0 (240822)
Warranty certificateVeroval
®
BPU 26
IFU_9251060_240822.indd 2
IFU_9251060_240822.indd 2
25.08.22 15:09
25.08.22 15:09

Related product manuals