HARVEY Hydroclave MC10 - User Manual

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
BARNSTEAD THERMOLYNE CORPORATION
Harvey
®
Hydroclave MC8
Hydroclave MC10
Steam Sterilizer
(With Optional Non-Recirculating System)
OWNER AND OPERATOR MANUAL

Table of Contents

Questions and Answers

  • S
    spadillaAug 6, 2025
    What to do if the load does not dry in the drying cycle of HARVEY Hydroclave MC10?
    • C
      Christopher HuberAug 6, 2025
      If your load is not drying in the drying cycle of your HARVEY Laboratory Equipment, consider these potential causes and solutions: * **Inadequate drying time:** Try increasing the drying time. * **Delayed door opening:** Ensure you open the sterilizer door immediately after the sterilize cycle finishes. * **Excessively wet load:** Check the load size, ensure items aren't touching chamber walls, and verify proper loading according to the 'Load Chart' and 'Loading' procedures. If using the open door drying method, keep the door open approximately 1/4" (push the latch down while the door is open).
  • S
    Shannon CarrilloAug 22, 2025
    Why are wrapped loads wet after sterilize cycle in HARVEY Hydroclave MC10?
    • C
      Christy BeckAug 22, 2025
      If wrapped loads are wet after the sterilize cycle in your HARVEY Laboratory Equipment, it might be due to improper loading. Make sure the wrap isn't touching the chamber walls and, if you have a 10" chamber, avoid placing wrapped goods on the bottom of the tray rack.

Summary

Operator Maintenance

Daily Maintenance

Daily checks and cleaning tasks for the sterilizer.

Weekly Maintenance

Weekly cleaning procedures for the chamber, racks, and trays.

Safety Information

Alert Signals

Explains warning, caution, and note symbols used in the manual.

Operation

Diagnostic Messages

Troubleshooting

Maintenance Schedule: Every 6 Months

Installation

Leveling Sterilizer Cabinet

Instructions for leveling the sterilizer cabinet side-to-side and front-to-back.

Electrical Connection

Guidelines for connecting the sterilizer to a power source.

HARVEY Hydroclave MC10 Specifications

General IconGeneral
Chamber Volume10 Liters
Water Level SensorYes
Material (Chamber)Stainless Steel
DisplayDigital LCD
Safety FeaturesOver-Temperature Protection
Power Supply50/60Hz
MaterialStainless Steel