Page
14
Example
6
TIME/CM-To-Frequency
Correlation
Eleven
vertical
and
nine
horizontal
lines
on
the
graticule,
spaced
1
centimeter
apart,
permit
the
measurement
of
displayed
waveforms.
The
short
markers
on
the
centerlines
are
spaced
2
millimeters
apart.
Use
the
following
formula
to
determine
the
frequency
of
a
waveform
displayed
on
the
CRT.
TIME/CM
switch*
setting
in
seconds
per
cm.
1
-
MAGNIFIER
Centimeters
for
1
Period
of
X
cycle
of
unknown
=
unknown
frequency
frequency.
pare
sie
and,
Period
~
Frequency
EXAMPLE:
If
one
cycle
of
a
waveform
measures
2
cm
on
the
graticule,
with
the
TIME/CM
switch
at
1
mS
and
the
X5
switch
pulled
out,
then
sey]
7
1
001°"
X
5
X
2
=
.0004,
and
0004
=
2500
Hz.
NOTES:
“The
VARIABLE
—X5
control
must
be
fully
clockwise
to
use
the
calibrated
TIME/CM
switch
settings
in
this
formula.
**1
millisecond
=
.001
second;
1
microsecond
=
.000001
second.
When
the
X5
switch
is
pulled
out,
the
sweep
travels
5
times
further.
(theoretically),
which
makes
the
sweep
time
shorten
by
a
factor
of
5.