VERTICAL
AMPLIFIER
BALANCE
NOTE:
During
the
following
steps,
you
may
experience
the
rare
situation
that
the
vertical
amplifier
of
one
channel
will
not
balance
properly.
If
this
happens,
interchange
transistors
Q101
and
Q102
of
the
channel
that
does
not
balance.
The
transistor
leads
are
not
soldered
to
the
socket
pins.
Channel
Y1:
Perform
the
following
numbered
steps 1-11.
Adjust
only
the
controls
associated
with
Channel
Y1.
1.
Adjust
the
POSITION
control
to
place
the
trace
on
the
horizontal
centerline.
2.
Turn
the
VARIABLE
control
fully
counterclockwise.
The
trace
may
move
off
the
screen.
3.
Adjust
the
POSITION
control
to
return
the
trace
to
the
center
horizontal
graticule
line.
4.
Turn
the
VARIABLE
fully
clockwise
(CAL).
5.
Slowly
adjust
the
DC
BAL
control
(a
screwdriver
adjustment
through
the
front
panel)
to
return
the
trace
to
the
center
horizontal
graticule
line.
6.
Repeat
steps
2
through
5
until
the
trace
does
not
move
when
you
turn
the
VARIABLE
control.
Leave
the
VARIABLE
control
in
the
fully
clockwise
position
(CAL).
7.
Turnthe
VOLTS/CM
switch
to
the
10mV
position.
8.
Adjust
the
BAL
control
[R124,
on
the
vertical
circuit
board
(see
Figure
16)]to
return
the
trace
to
the
center
horizontal
graticule
line.
9.
Turn
the
VOLTS/CM
switch
back
and
forth
between
the
10mV
position
and
the
50mV
position
and
adjust
the
BAL
control
so
the
trace
does
not
move
when
the
switch
is
turned
from
one
position
to
the
other.
Page
27
10.
Repeat
steps
2
through
5.
11.
Turn
the
Y1
POSITION
control
fully
counterclockwise
(OFF).
Then
turn
the
Y2
POSITION
control
clockwise
and
center
the
trace
on
the
screen.
Channel
Y2:
Perform
the
previous
numbered
steps
1-10.
Adjust
only
the
controls
associated
with
Channel
Y2.
TRIGGER
1.
Set
the
Y1
and
Y2
INPUT
switches
to
GND.
2.
Adjust
the
Y1
and
Y2
POSITION
controls
to
place
both
traces
on
the
horizontal
centerline.
3.
Adjust
your
voltmeter
to
measure
—3.0
volts
DC.
4.
Connect
the
common
voltmeter
lead
to
the
Oscilloscope
chassis.
5.
Measure
the
voltage
at
TP14
(test
point
14)
on
the
vertical
circuit
board.
(See
Figure
16
in
the
“Illustration
Booklet.’’)
Adjust
the
Y1
ZERO
control
(R149-1,
on
the
vertical
circuit
board)
for
zero
volts.
6.
Measure
the
voltage
at
TP15
on
the
vertical
circuit
board.
Adjust
the
Y2
ZERO
control
(R149-2,
on
the
vertical
circuit
board)
for
zero
volts.
7.
Disconnect
the
voltmeter.