10.
11.
12.
13.
14,
15.
16.
17.
18.
19.
Readjust
the
SWEEP
LENGTH
control
(R255,
on
the
horizontal
circuit
board)
so
the
trace
is
10
cm
long.
Turn
the
HOR!Z
POS
control
counterclockwise
to
move
the
right-hand
end
of
the
trace
1
cm
to
the
left.
Adjust
the
SWEEP
LENGTH
control
(R255,
on
the
horizontal
circuit
board)
to
move
the
right-hand
end
of
the
trace
back
to
the
right
1
cm.
Readjust
the
HORIZ
POS
control
so
the
left-hand
edge
of
the
trace
starts
at
the
left-hand
graticule
line.
Pull
out
the
VARIABLE
X5
switch.
Adjust
the
X5
CAL
control
(R278,
on
the
horizontal
circuit
board)
and
the
HORIZ
POS
control
for
exactly
2
cycles
in
10
centimeters
as
shown
in
Figure
19.
When
you
make
this
adjustment,
place
the
leading
edge
of
the
first
cycle
on
the
left-hand
graticule
line
and
the
leading
edge
of
the
second
cycle
on
the
center
graticule
line.
Push
in
on
the
VARIABLE
X5
switch.
Place
the
Y1
INPUT
switch
in
the
GND
position.
Adjust
the
RAMP
ZERO
control
(R248,
on
the
horizontal
circuit
board)
until
the
bright
spot
on
the
left
end
of
the
trace
disappears.
(This
is
not
a
critical
adjustment.)
Place
the
INPUT
switch
to
AC.
Refer
to
Figure
17
in
the
‘‘IIlustration
Booklet,”
and
carefully
readjust
the
X
CAL
control
(R268,
on
the
horizontal
circuit
board)
and
the
HORIZ
POS
control
for
exactly
10
cycles
in
10
centimeters
(1
cycle
per
centimeter)
as
shown
in
Figure
18.
Use
the
leading
edge
of
the
second
and
tenth
cycle
for
this
adjustment.
Turn
the
TIME/CM
switch
to
1
uS.
Connect
a
1
MHz
square
wave
signal
to
the
Y1
INPUT
connector.
Use
a
plastic
alignment
tool
and
adjust
trimmer
capacitor
C229
(located
on
the
TIME/CM
switch)
for
1
cycle
per
centimeter.
Disconnect
the
square
wave
signal.
[
I~
Wittttitiiiiiiii
HY
itty
hth
tae
mi
VERTICAL
‘ines
one:
eens
nee
ee
anne
een
mene
one
eae
Page
29
au
Sbsiiaians
es
ae
"ss
aaa
se
aaa
LEADING
EDGE
ari
H
x
i
LEADING
EDGE
or.
SECOND
Crue
H
TESTS
bg
soto
i.
sbiacoseil
PORTION
OF}
fo}
avses
TRACE
WILL
=...
sae
BE
DIM.
.
;
wrnnontannnnals
Se
CER
AE:
SRP
EON
AO,
OS
1
CYCLE/cm
Figure
18
deeee
ccndanee
Lucite
Figure19
Turn
the
Y1
VOLTS/CM
switch
to
the
200mV
Position.
Turn
the
TIME/CM
switch
to
the
X-Y
position.
Turn
the
TRIG
switch
to
Y2,
+.
Adjust
the
Y1
and
Y2
POSITION
controls
to
center
the
spot
on
the
graticule.
Adjust
the
HORIZ
POS
control
so
the
Y1
POSITION
control
will
move
the
spot
off
the
screen
in
both
directions.
Set
the
Y1
INPUT
switch
to
AC.
Connect
the
Y1
test
cable
inner
lead
to
the
1V
P-P
connector
on
the
front panel.
NOTE:
A
square
wave
voltage
calibrator
may
be
used
instead
of
the
oscilloscope
1V
P-P
output.