To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
of
a
variable
DC
power
supply.
C
supply
voltage
is
o
tained
by
using
t¥6
AC
sources
(one
fixed
na
one
tapped),
two
silicon
rectifiers
(CR-1
and
CR-2)
and
a
charging
capacitor
(C-3).
DC
po-
tentials
are
applied
to
the
voltage
regulating
tube
under
test
through
a
current
limiting
re-
sistor
(R-7)
which
also
serves
as
the
regulating
resistor.
A
voltmeter
circuit
(R-5,
R-6
and
the
meter)
indicates
the
firing
voltage
as
the
applied
potential
is
varied
over
the
range
in
which
the
tube
should
fire.
The
point
at
which
the
tube
fires
is
indicated
by
a
sudden
downward
deflection
of
the
meter
pointer.
This
occurs
because
it
is
the
property
of
a
voltage
regulator
tube
to
main-
tain
a
voltage
across
its
elements
in
the
order
of
five
to
thirty
voits
below
its
firing
potential.
After
the
VR
tube
has
fired,
its
value
as
a
reg-
ulator
can
be
determined
by
varying
the
applied
voltage
between
the
limits
specified
on
the
roll
chart.
Between
these
limits
a
variationfrom
the
regulating
point
will
be
an
indication
of
the
tube's
THYRATRON
TEST
CIRCUIT
for
their
grid
c
tinn
eanahiliti
Veblen
Ww
Ap
AN
a
se
vd
nected
into
the
gram.
r
.
The
tube
tob
est
circuit,
as
s
x
&
o}
2
5
ct
De.
®
ta
ok
&
For
this
test
the
meter
is
connected
inthe
plate
circuit
as
a
milliammeter.
The
tube
is
held
to
cutoff
by
means
ofa
high
bias
applied
to
the
grid.
When
the
bias
is
decreased
so
that
the
grid
be-
comes
less
negative
with
respect
to
the
cathode,
a
point
will
be
reached
when
the
tube
will
sudden-
ly
conduct
and
give
an
indication
on
the
meter.
The
value
of
BIAS
setting
at
which
this
occurs
should
fall
within
the
limits
specified
in
the
REMARKS
column
of
the
roll
chart,
For
ex.
Ve
ave
See
eS
USS
sl
25)
ample,
a
tube
having
"Thyra;
Fire
15-0
Bias"
tha
DEMADYS
anlumn
should
fire
hetwean
in
LID
ORVIUUVISARUEN
UULUILIE
OLIUUL
£24E0U
WUL
WU
a
BIAS
setting
of
15
and
0.
After
the
tube
has
fired,
the
pointer
should
fall
within
the
specified
portion
of
the
scale
for
the
tube
to
be
considered
good.
-_
.
eS
HEATHKIT
—S
[RZ
67
2V.
2K
10W
fra
(eri)
fires
at
110
en
let
us
assume
that
it
will
i
ating
potential
of
approximately
105
volts,
with
the
following
in
the
REMARKS
column:
"4.0
V
Reg,
Fil
5
to
70
V."'
Therefore,
when
the
FILAMENT
switch
is
rotated
from
5
to
70,
the
meter,
which
now
acts
as
a
voltmeter
across
the
VR
tube,
should
not
vary
more
than
4
volts.
If
the
variation
is
more,
the
VRtube
should
be
considered
faulty
and
replacement
is
recom-
mended.
ma
ntain
fur]
feat
a1
eo
ike
OR
iR3
620n
oon
>——-5)
|
Be
_|
|]
|
|
oc
oc
BIAS
PLATE
SUPPLY
\
SUPPLY