8
When accurate setting for C=0.010 and C=0.100, the shift x 10 blinks to
prompt extension ratio due to the screen display adopt X10 mode.
Ⅳ.Failure judgement
When the reading is incorrect or unstable, please check the meter and electrode.
1. Under resistivity measurement condition, remove the white wire from the wire
terminal and check the resistivity reading, if the reading is 18.23MΩ•cm and
stable, the meter is good. The problem can be initially identified from the
electrode installation.
2. Under conductivity measurement condition, remove the white wire from the
wire terminal and check the conductivity reading, if the reading is zero and
stable, the meter is good. The problem can be initially identified from the
electrode installation.
1) Connect the white and green terminals on the short-circuit meter and check the
resistivity reading, if the reading is 0.0M•cm,the meter is good. The problem
can be initially identified from the electrode installation.
2) Meter mode or transmitter mode of (4~20)mA judgment.
Before the judgment please makes sure there is no wire connection on the terminal.
C=0.010, electrode constant value is sensor’ s label value X 10(range: 0.0801-
0.1199)
C=0.100, electrode constant value is sensor’s label value X 10(range:0.801-
1.199)
C=1.000 electrode constant value is sensor’ s label value X 10(range:
0.801-1.199)
Digital
input
Enter
Enter
C=10.00 electrode constant value is sensor’ s label value X
10(range:8.01-11.99)
Digital
shifter