2 - Maxima Advanced Ultrasonic Cleaner MA210
GENERAL OPERATIONSSAFETY PRECAUTIONS
• Do not turn on the unit without water in the tank. Dry running over 30 seconds will damage
the unit or severely reduce the life of the unit.
• Do not place instruments directly on the base of the tank. Placing instruments directly on the base
of the tank will result in damage to the unit. Place instruments into the basket for proper use.
• Do not run the unit without water in the tank.
• Normal cleaning time is 5 minutes only per layer of instruments or cassette, no more than
10-15 minutes per batch. Take a break for 1/3 to 1/2 of the running time if longer running time is
needed.
• Do not cause water to get inside the housing through vents or other places. Do not wash it,
only wipe it.
• Do not spray disinfectant towards the unit otherwise it may cause corrosion internally. Only
wipe it.
• Keep the lid closed during running. Beeping sound will be heard if the lid is opened during
running.
• Do not run the unit without the lid on. The lid helps reduce aerosols and reduce noise.
• Avoid cleaning burs and files in the main tank directly otherwise they may get into the drain-
age valve and cause damage. Use a beaker or cleaning chamber. See accessories.
• Do not fill the tank above the Max line to avoid overspill.
• Do not use solution containing abrasive substances or strong corrosive chemical solution not
recommended by the manufacturer or the supplier.
• Do not use outdoors.
• To reduce the risk of electrical shock, do not put the appliance in water or other liquid. Do not
place or store appliance where it can fall or be pulled into a tub or sink.
• Do not run the unit for an extended time or continuously. If the unit has been running for 30
minutes, it is recommended to stop the unit for about 10 minutes.
• Do not use components unapproved by the manufacturer.
• Place the device on a dry and flat surface for operation.
• If any abnormalities are found, stop using immediately and contact Henry Schein® Handpiece
& Small Equipment Solutions.
• This device should only be operated by a professional.
• Never operate the device unattended.
• Disconnect the power cord from the unit before cleaning the device, if malfunction occurs, if
the device is not going to be used for prolonged period, and after each use (recommended).
• Follow the manual to operate the device.
• Never disassemble the unit. All servicing of this unit is to be conducted by a qualified service
personnel.
• When removing the power cord from the socket, grab the power plug, not the cord.
• To protect the power cord from damage, do not cause it to get caught by things such as a
cupboard door or a chair leg; do not drag across a hot surface.
• If there is damage to the power plug, cord, housing, or other parts of the device, do not use the
device.
• The installation of an earth leakage circuit breaker with a rated tripping current of no more
than 30mA provides further protection against an electrical shock. The installation should only
be carried out by a trained electrician.
Keep away from children!
• Please store the ultrasonic cleaner where it is not reachable by children.
• Danger to children! Danger for death through suffocation! Keep the packaging material away
from children.
• This appliance shall not be used by children. Keep the appliance and its cord out of reach of
children.
To prevent fire hazards, please observe the following:
• Never block the vents on the device. Keep the vents free from lint, hair, and other materials.