Maxima PowerClean 210 - 4
CONTROL PANEL
1 63 7
2
5
4 8
1. Degas dial. Turn it clockwise or counterclockwise to select the degas timer setting, either
60, 120 or 180 seconds. This setting determines the automatic and manual degas time. Push
in the dial to start degassing.
2. Warning. Stop operations and see troubleshooting guide (pg 7).
3.
Open lid warning. The lid should be closed during operation to prevent the spread of
aerosols. The warning symbol helps to standardize operations. The open-lid warning symbol
is illuminated if the unit is not in cycle. Additional warning beeps are sounded if the unit is in
cycle.
4. 60, 120 or 180 Degas timer display. It displays the degas timer setting and how much
degas time remains in seconds (s).
5.
Degas status. ashes, indicating degas is underway.
6. 05:00. Ultrasonic time display. It displays the timer setting and how much time remains in in
the ultrasonic cleaner operation.
7.
Ultrasonic cleaning status. illuminated, normal cleaning is under way.
8. Set time dial/start and stop. Rotating either clockwise or counterclockwise can set the
timer to 6 settings: 5:00 – 7:30 – 10:00 – 12:30 – 15:00 – 20:00 mins. Once the timer is set,
push in the dial to start.
SETUP
1. Place your ultrasonic cleaner in a desired location on a dry fl at surface next to a sink.
2. Connect the hose. Connect one end of the drain hose to the drain valve and place the other
end in the sink. The drain valve is located behind the unit on the right side.
3. Plug In the unit. Attach the power cord into the back of the unit. Ensure that it is properly
attached and that the cord is undamaged.
4. Turn unit on. The power switch to turn the unit on is located in the back of the unit on the left
side just above the power socket.
OPERATION
Do not turn on the unit without water in the tank. Dry running will damage the unit.
Do not place instruments directly on the base of the tank. Place instruments into the
basket for proper use.
Avoid cleaning burs directly in the main tank. Use beakers or cleaning chamber.