EasyManua.ls Logo

Heraeus HC 0057 - Suggestion Regarding Fault Signal; Lack of Water Humidification System

Default Icon
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HC 0020
-
HC
0057
HC 2020
-
HC
2057
SERVICE MANUAL
2.
Suggestion regarding
fault signal "lack of water
humidification system"
(Concerning HG 0/2 only). Should this signal continue although
the supply tank holds sufficient water (at least half full) and the pump
trouble has been corrected ace. to operating instructions, the solenoid
valve 11-Y54 might be responsible for the malfunction; the reason
being that dirt particles on the valve seat impede proper closing of the val-
ve.
Consequently, the pump takes in air and the water circulation can't be
started.
To remove dirt, dismantle solenoid valve and clean the valve seat.
Then start humidification system again.