The lens is very delicate.
It is very important to protect the lens from dirt and damage.
Use a clean, soft cloth to clean the surface of the device and LCD. Do not use
solvents or immerse the device into water or other liquids.
Always keep the thermometer within the storage temperature range (- 20℃to 55
℃or – 4℉to 131℉) and humidity range (≤93% non-condensing).
It is recommended to store the thermometer in a dry location free from dust. Do
not expose the thermometer to direct sunlight, high temperature/ humidity or any
extreme environment, otherwise the function will be reduced.
When the ambient temperature of the thermometer changes too much, such as
moving the thermometer from one place of lower temperature to another place of
higher temperature, allow the thermometer to remain in a room for 30 minutes where
the temperature is between15℃ to 40℃.
Used batteries should not be disposed of in the household rubbish. Used Batteries
should be deposited at a collection point.
At the end of its life, the appliance should not be disposed of in household rubbish.
Enquire about the options for environment-friendly and appropriate disposal. Take
local regulations into account.
Our company warrants Non Contact Infrared Body Thermometer at the time of its
original purchase and for the subsequence time period of one year.
The warranty does not cover the followings:
When asked to provide warranty service, you must have a purchase date and
purchase stamp dealers (including dealers name and address) of the warranty card.
Be sure to ask the dealer to purchase this product signature on the warranty card.
When asked to provide warranty service, please put the product to get our
distribution points for repair. Products outside the warranty expires, will be charged
accordingly.
The device series number label is torn off or cannot be recognized.
Damage to the device resulting from misconnection with other devices.
Damage to the device resulting from accidents.
Changes performed by users without the prior written authorization of the company.
Batteries and packaging are not covered under warranty
1. If you have any problems with this device, such as setting up, maintaining or using,
please contact with SERVICE PERSONNEL of HeTaiDa Technology Co., Ltd. Don’t open
or repair the device by yourself.
2. Please report to HeTaiDa Technology Co., Ltd. if any unexpected operation or
events occur.
3.
4. The patient is an intended operator. The patient can measure and change battery.
Under normal circumstances and maintain the device and its accessories according
to the user manual.
1) This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS;
This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be
affected by portable and mobile RF communications equipment.
2)* Caution: Do not use a mobile phone or other devices that emit electromagnetic
�ields, near the unit. This may result in incorrect operation of the unit.
3) *Caution: This unit has been thoroughly tested and inspected to assure proper
performance and operation!
4) * Caution: this machine should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, this machine should be
observed to verify normal operation in the con�iguration in which it will be used.
Note:
Guidance and manufacture’s declaration – electromagnetic emission
Guidance and manufacture’s declaration – electromagnetic immunity
Anti-interference detection IEC 60601 test level Compliance level
Emission test Compliance
The Infrared Body Thermometer is intended for use in the electromagnetic environment speci�ied below. The
customer of the user of theInfrared Body Thermometer should assure that it is used in such an environment.
The Infrared Body Thermometeris intended for use in the electromagnetic environment speci�ied below. The
customer or the user of Infrared Body Thermometer should assure that it is used in such an environment.
RF emissions
CISPR 11
RF emission CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage �luctuations/
�licker emissions
IEC 61000-3-3
Group 1
Class B
Not applicable
Not applicable
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
interruption
IEC 61000-4-11
Power frequency
(50Hz/60Hz)
magnetic �ield IEC
61000-4-8
250 (50Hz)/300 (60Hz) cycles
for >95% UT (sync angle
(degrees):0)
30A/m 30A/m
Voltage dips
IEC 61000-4-11
0.5 cycles for > 95% (sync
angle (degrees):0, 45, 90,
135, 180,225, 270, 315)
1 cycles for >95% UT (sync
angle (degrees):0)
25 (50Hz)/30 (60Hz) cycles
for 30% U T (sync angle
(degrees):0)
The input a.c. power ports: ±2 KV
The input d.c. power ports: ±2 KV
Signal input/output ports: ±1 KV
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Contact: +8 KV
Air: ±2, ±4, ±8, ±15 KV
Contact: +8 KV
Air: ±2, ±4, ±8, ±15 KV
Not applicable
Not applicable
NOTE UT is the a.c. mains voltage prior to application of the test level.