T
Terry LawsonAug 4, 2025
What to do if Heyl Testomat 808 shows error E4?
- WWilliam ThorntonAug 4, 2025
If your Heyl Analytical Instruments shows error E4, it could be due to insufficient LED power or light. First, check if the water is turbid, as strong turbidity can negatively influence the measurement. Also, inspect the measuring chamber or sight glass for soiling. The sensor or LED might be damaged. Ensure the water level isn't too low. Verify that the measuring chamber and the LED fitting are installed correctly, as they can loosen during transport, disrupting the measurement. If needed, perform a visual adjustment as described in the maintenance instructions for the Testomat® 808.