6. Operating the device
Nebulization directly from the nebuliser chamber
After the device is assembled use a sterilized nebulizer system and ensure that both,
air filter (33) and bacteria filter (23), are properly installed.
• Medication in doses between 5ml and 10ml can be nebulized.
• Remove the lid from the nebulization chamber (11) by turning it slightly back and forth
and pulling at the same time.
• Plug the conical medication cup (38) on the nebulization chamber lid (36) make sure
the water is between the minimum and maximum levels.
Apply desired medication (max 10ml into the disposable medication cup (38).
• Connect the aerosol tube (15) and ventilation tube (16) to the chamber lid.
Turn on the device using the ON/OFF switch which is located on the left side of the
control panel. The nebulization process will start.
Note: The distance between the device and the ground should not exceed 60 cm.
16
36
38
39
11
11.
Nebulizer Chamber
16. Ventilation tube
36. Chamber lid
38. Medical cup
39. Transducer