Respiration Therapy Device TREND 300
Functions of TREND 300
WAKE TIME..........06:00
EXIT MENU!
Press Input key to leave the programming mode.
During standard operation the display indicates
date and time:
SA 28.FEB 12:42
The alarm clock can be activated or deactivated by the
keys and .
With the key the alarm clock can be switched on. For
controlling the programmed alarm time is indicated for a short
time in the display. After this a bell symbol shows that
the alarm clock is active.
SA 28.FEB
12:42
With the key
the alarm clock can be switched off (e.g. on
Sunday). In this case the bell symbol is not indicated.
NOTE:
If there is no key pressed during programming for
30 seconds the programming mode will be left
automatically.
All functions are programmed in the same way:
Press for 1 s, and choose the line requested with the
keys and :
By pressing the key the following options can be
chosen:
WAKE TIME WAKE HOUR, WAKE MINUTE
TIME HOUR, MINUTE
DATE YEAR, MONTH, DAY, WEEKDAY
RAMP 1. mask test time: 0...90 s (in 5 s-steps)
2. softstart time: 0...60 min (in 1 min-steps)
AUTOMATIC ON, OFF
I-SET 1. PIN INPUT! following selection after:
2. 3,0 ... 18,0 hPa (in 0,5 hPa-steps)
E-SET 3,0 ... 18,0 hPa (in 0,5 hPa-steps)
I-SLOPE 0,0 ... 1,5 s (in 0,1 s-steps)
E-SLOPE 0,0 ... 1,5 s (in 0,1 s-steps)
TRIGGER 1 ... 5
MODE BILEV-S (only indicate, not selection possibility)
BRIGHTNESS 100%, 50%
COUNTER POWER, BLOWER, FILTER, THERAPY (time indicate in h)
EXIT MENU!
If you want to finish the programming, choose the menu
line
EXIT MENU! and press key .