MAN-04008-001 Rev. 001 Page 20 of 27
ThinPrep 2000 S
y
stem
Three lab CTs have read the same
ThinPrep 2000 slide from a vial
Three or two CTs
had ASC-H+
Three or two CTs
had <ASC-H
ThinPrep 5000
Processor
Three lab CTs have
read the same
ThinPrep 5000 slide
from a vial
Three or two CTs
had ASC-H+
194 34 242
Three or two CTs
had <ASC-H
26 1006 1032
220 1040 1260
Agreement among Laboratory Cytotechnologists/Pathologists
The following tables indicate the extent to which the laboratory cytotechnologists / pathologists at a given site
agreed amongst themselves on the diagnosis, comparing the ThinPrep 5000 processor to the ThinPrep 2000
system. Tables are provided for ASC-US+ and ASC-H+.
In Table 24 for ASC-H+, the number of specimens is shown for which various levels of agreement among
the CTs occurred. Either all three CTs rated the slide as positive (ASC-H+), two out of three rated it
positive, one out of three, or none of them.
Table 24: Laboratory Cytotechnologist/Pathologist Agreement, All Results, ASC-H+
ThinPrep 2000 System
Three lab CTs have read the same ThinPrep 2000 slide from a vial
ASC-H+
Three CTs
had
ASC-H+
Two CTs had ASC-H+
& one had <ASC-H
One CT had ASC-H+
& two had <ASC-H
Three
CTs had
<ASC-H
Totals
ThinPrep
5000
Processor
Three lab
CTs have
read the
same
ThinPrep
5000 slide
from a vial
Three CTs had
ASC-H+
111 21
6
0 138
Two CTs had ASC-H+
and one had <ASC-H
32 30
21
7 90
One CT had ASC-H+
and two had <ASC-H
7 9
43 28 87
Three CTs had
<ASC-H
2 8
37
898 945
Totals
152 68 107 933 1260
ASC-H+
Totals
Totals