Hospitex Diagnostics EOS Bravo and EOS Bravo Plus Service Manual
34
16) connect all the connectors of the cabling on the dispositives present on the instrument
according to the following scheme:
- MX254 6 poles with the note MDS on the motor of the diluter using a motor protection diode
module
- MX254 3 poles with the note SDS on the optical limit switch of the diluter
- MX254 2 poles with the note EV1 on the electrovalve of the diluter
- MX254 6 poles with the note MS on the motor of the sampling needle (through the rotating
connection) using a motor protection diode module
- MX254 3 poles with the note SSN on the optical limit switch of the sampling needle (through
the rotating connection)
- MS396 2 poles with the note R2 on the heating reel (through the rotating connection)
- MX254 6 poles with the note MRP on the motor of the reagent plate using a motor
protection diode module
- MX254 3 poles with the note SRP of the optical limit switch of the reagent plate
- MX254 6 poles with the note MIP on the motor of the sample plate using a motor protection
diode module
- MX254 3 poles with the note SIP of the optical limit switch of the sample plate
- MX396 2 poles with the note R1 on the heating reel of the reaction plate