- MX254 2 poles with the note ST2 on the temperature sensor of the reaction plate
- Faston couple, coming out from the cabling together with the cables R1 and ST2, on the
security thermostate of the reaction plate
- MX254 6 poles with the note MOP on the motor of the reaction plate using a motor
protection diode module
- MX254 3 poles with the note SOP on the optical limit switch of the reaction plate
- MX254 6 poles with the note MAN on the motor of the reading needle using a motor
protection diode module
- MX254 3 poles with the note SAN on the optical limit switch of the aspiration needle
- MX254 6 poles with the note MPM on the motor of the peristaltic pump using a motor
protection diode module
- MX254 6 poles with the note MFS on the filter wheel motor using a motor protection diode
module
- MX254 3 poles with the note SFS on the optical limit switch of the filter wheel
- MX254 6 poles with the note FD1 on the cable coming form the preamplifier board
- MX254 2 poles with the note ST1 on the temperature sensor of the flow cell
- MX396 2 poles with the note PT1 on the Peltier cell of the flow cell thermostation
- MX396 2 poles with the note L1 on the optical lamp
- Faston couple coming out from the cabling together with the cables L1 and PT1 on the
security thermostate of the flow cell
- Flat 16 poles of the serial port on the cpu board to the serial port of the PC104 board
17) Take everything in high towards the electro valve, the diluter piston rotating manually the
command screw connected to the motor tree.
18) turn on the instrument and make sure that within 1-2 seconds the printer executes an empty
course of the head.