M9 Transfer Instructions For Use (M999-19, Version 10, 2021-12) 35
12. SPECIFICATIONS
12.5 ELECTRIC ACTUATOR SYSTEM
Manufacturer: LINAK
Power supply voltage: 220-240 V AC
Power supply frequency: 50-60 Hz
Operating voltage: 24 V DC
Maximum input current: 2.0 A
Duty cycle: 10%, do not operate the stretcher for more than 2-minutes over a 20 minute
period.
12.6 TRANSPORT & STORAGE ENVIRONMENT
Ambient temperature range: +5°C to +40°C
Relative humidity range: 30% to 75%
Atmospheric pressure range: 80 kPa to 106 kPa (Rated to operate at an altitude ≤ 2000m).
12.7 OPERATING ENVIRONMENT
Ambient temperature range: +5°C to +40°C
Relative humidity range: 30% to 75%
Atmospheric pressure range: 80 kPa to 106 kPa (Rated to operate at an altitude ≤ 2000m).
12.8 SOUND PRESSURE LEVEL
Sound pressure level: 51.0 dBA (MAX)