Performance
Tests
Model
3336A/B/C
DC
VOLTMETER
IMPEDENCE
MATCHING
PADS
(AS
REQUIRED)
OV,
73
OHM
THERMAL
CONVERTER
IMPEDANCE
MATCHINS
PADS
150
TO 75
GHM
o_o
699
TO
75
OHM
o_o
R1
BALANCED
754
BALANCED
750
BALANCED
750
IMPEDANCE
UNBALANCED
IMPEDANCE
UNBALANCED
IMPEDANCE
UNBALANCED
R1
Re
r
:
Bo
Figure
4-3.
Amplitude
Flatness
Equipment
Set
Up.
124
AND
135
TO
75
OHM
c.
Allow
time
for
the
thermal
converter
to
settle
before
taking
any
measurements.
The
accuracy
of
this
test
is
enhanced
if
the
thermal
converter
is
protected
from
drafts
and
other
sources
of
temperature
change.
Measure
and
record
the
thermal
converter’s
output
with
the
3336
programmed
at
10
kHz.
Use
extreme
caution
when
making
the
following
measurements
at
the
75
ohm
output.
The
converter
voltage
will
approach
maximum,
and
the
converter
will
be
destroyed
if
care
is
not
exercised.
d.
Measure
the
thermal
converter
voltage
for
each
output
at
the
frequencies
listed
above.
At
each
frequency
setting,
adjust
the
amplitude
of
the
3336
in
.01
dBm
steps
until
the
con-
verter
voltage
is
equal
to
the
reference
voltage
recorded
at
10
kHz.
When
the
converter
level
matches
the
reference
level,
record
the
amplitude
setting
of
the
3336.
Do
not
adjust
the
amplitude
of
the
3336
in
increments
larger
than
.01
dBm.
@
4-16