Section
1
Model 3455A
Table
1-3.
Recommended Test
Equipment.
Instrument
Critical Specification Recommended
Model Use
DC
Voltage Standard Voltaga: 10
mV to 1000
V
Accuracy: i .005%
Syetron Donner
Model
Ml
07
PAT
AC
Calibrator Frequency: 30 Hz to 100
kHz
Output
Level:
100
mV
to 1000 V
Accuracy: i .1
%
Voltage
Stability
<6
moi.)
t .02%
-hp- Model
745A
AC
Calibrator
hp- Model 746A
High Voltage Amplifier
PAT
Test Oscillator Frequency: 10 250 kHz
Output: 3 V
rms into
50 n
Frequency Response t
.25%
-hp-
Model
6S2A
Test Oscillator
P
Resistance Decade
Resistance:
100 n to 10 MD
Accuracy:
t
.004%
Gen
Rad Model i
GR 1433-Z
Decade
Resistor
PAT
DC Null
Voltmeter Voltage Range.
1
siV
to
10
V
-hp- Modal
419A
PAT
Reference
Divider Division Ratio Accuracy t .001%
Output Voltage
Range
-
1 V 10 1 kV
Fluke Model
750A
Refereitce Divider
PA
DC
Transfer Standard Output
Voltages:
1 V, 1 .018 V.
1.019V.
10
V
Accuracy: t
5
ppm
Stability: t .001% <30 daysl
Fluke Model
731A
DC
Transfer
Starrdard
PA
Electronic Counter 50 Hz to 60 Hz
-hp- Model
S300A/S302A
Measuring System
P
ResKtance
Standard Resistance: 1 kfl
Accuracy :
± .0005%
Resistance:
100 K
Accuracy:
± .002%
Guildine Model
9330/1 K
or 9330A/1
K
Guildline
Model 9330/100
f
A
Bus System
Analyzer
HP-IB
Control Capability -hp- Model
59401
A
Bus System Analyzer
T
Calculator HP-IB
Control Capability must
serve
as
printer for 3455A
Output data.
-hp- Model
9825A OT
Oscilloscope Bandwidth: DC to 10 MHz
Sweep
Time:
0.1 ps to 1 sec/div
Sensitivity: 1 V/div
-hp- Model
180C/D
Oscilloscope with
IBOIAand
1821A
plug-in units
T
Digital Voltnteier Voltage Range; 10 mV to 1000 V
Resolution:
lOpV
-hp. Model
3490A PAT
Raiisiori Resistances:
1 kn * 10%
10 kn 1 0.1%
1 Mn to.1%
-hp. Part
No.
0684-1021
0698-4157
0698-6369
P
Signature Artalyzer
-hp- Model S004A
T
P Performance
Checks T
Troubleshootirtg
A
-
Adjustmems 0
» Operetors Check
1-8