Model 3455A
Section
IV
Table
4-4.
AC Accuracy
Tast
30
Hz
to
100 kHz
(Standard Model 3455A only).
Test
Frequency
Input
Level
34S6A
Range
24 Hour**
Test
Limits
90 Day**
Test Limits
30
Hz* 1 V 1 V
0.99920 to 1.00080 0.99900 to
1.00100
100 kHz 1 V
1
V 0.99S20
to
1.00480 0.99400
to
1.00600
30
Hz* 5V 10 V
4 9940
to
5.0060 4.9925 to 5.0076
100 kHz 5V
10 V 4.9720 to
S.0280
4.9650
to
5.0350
30
Hz* 10
V
10 V 9.9920 to
10.0080 9.9900
to
10.0100
20
kHz 10 V 10 V 9.9920 to
10.0080 9.9900 to 10.0100
100 kHz 10 V
10
V
9.9520
to
10.0480
9.9400
to
10.0600
30 Hz*
100
V
100 V 99.920 to 100.080
99.900
to
100.100
100 kHz 100 V 100 V 99.520 to
100.480 99.400 to 100.600
30 Hz* 1000 V 1000 V
998.00
to
1002.00 997.50
to
1002.50
10 kHz 1000 V 1000 V 998.00 to
1002.00 997.50to
1002.60
*
Fraquenciet below 300 Hz apply to ACV
Function only.
*
'These test limits do not
Include
the
temperature coefdcients that must be edded if
the
Instrument is operated outside of
the temperature ranpe over
which
the
24-
hour or
90-dav
specifications
apply
(see
Table
1-1).
Derive
6-month test limits
from AC Accuracy specifications listed in
Table 1-1.
verify the
3455A
ac
voltmeter accuracy for each
Test Frequency {Fast
ACV),
Input
Level and
3455A Range listed in
Table
4-6.
The 3455A
display
readings should be
within
the Test
Limits
given
in the table.
h. Set the AC
Calibrator for an output of 1
V,
10
kHz. Set the 3455A FUNCTION to
ACV
and
RANGE
to 1 V.
i. Record the 3455A reading:. .
V.
j.
Set the 3455A
FUNCTION to FAST
ACV.
Record
the 3435A reading: . .
V.
k. Set the 3455A
FUNCTION
to
ACV and RANGE
to
10 V. Set the AC Calibrator for an output of 5
V, 10
kHz.
l. Record the 34S5A reading:. .
V.
m. Set the
3455A FUNCTION to FAST
ACV.
Record the 3455A reading: . .
V.
n.
Disconnect
the
AC Calibrator from the 34SSA. Set
the 3455A FUNCTION to
ACV
and RANGE to
1 V.
0
. Set the Test Oscillator for an output of 1
V, 10
kHz. Connect the SO-ohm output of the Test
Oscillator,
terminated in a 50 ohm
load,
to
the
34SSA
front panel
INPUT.
p.
Adjust the Test Oscillator
level
controls for a
3455A reading as close as possible to the reading record-
ed in Step i. Set
the
Test
Oscillator’s meter switch
to ex-
panded scale and adjust the meter reference controls for
a zero reading on the Test Oscillator’s
meter.
Use the
Test Oscillator’s
level
controls to maintain
this zero
reading
whenever
the Test Oscillator frequency is
varied.
q.
1. Standard Model 3455A: Set the Test Oscillator
to 1 MHz (maintain reference
level
on meter of
Test Oscillator). The 3455A display reading
should be
within
the Test Limits
given
in
Table
4-5.
2. 3455A Option 001: Set the Test Oscillator
fre-
quency to
250 kHz (maintain reference level on
meter of Test Oscillator). The 3455A display
reading should be
between 0.99240 V
and
1.00760 V (24-hour
spec.)
or between 0.99190 V
and 1.00810 V (90-day spec.).
r. Set the 3455A FUNCTION to FAST
ACV.
Set the
Test Oscillator frequency to 10 kHz and adjust its out-
put
level
for the 3455A reading recorded in step
j.
Ad-
just the meter reference controls for a zero
reading
on
Table 4-5.
AC
Accuracy
Teete
100
kHz to 1
MHz IStandard Model 345SA only).
Test
Frequency
Input
Level
3455A
Range
24
Hour*
Test
Limits
90 Day*
Test
Limits
1 MHz 1 V
1 V
0.92400
to
1.07600 0.90900 to 1.09100
1 MHz 5V
10 V 4.4900 to
5.5100 4J900 to 5.6100
350 kHz 5V
10
V
4.7600 to 5.2400
4.70IX)
to
5.3000
*Thes« lest limits
do
not
include the temperature coefficients
that must be added if
the instrument is operated
outside of the temperature
range over which the
24-hour
or
90-day
specifications apply (see Table
1-11. Derive 6-month test
limits from AC
Accuracy specifications
listed
in
Table
1-1
.
4-7