EasyManuals Logo

HP 4145A Operating And Service Manual

HP 4145A
269 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
SECTION I Model 4145A
Table l-l. Wecifications (Sheet 2 of 8)
I
Current Range
I
Resolution
I
Accuracy
+lOOmA lOOnA
+lOmA
1 OpA
0.3%+(0.1+0.2*Vo/100)%
I
+1000l~A
I
II-IA
I
I
+1 OOuA
I
1 OOnA
I
I
*lOuA
I
1 OnA
I
111 0.5%+(0.1+0.2*Vo/100)%
111 1%+(0.1+0.2*Vo/100)%+5pA
Max. Voltage
2ov (>SOmA)
4ov (>2OmA)
1oov (62OmA)
Setting Resolution :
Voltage, 4*1/2 digits (1mV max.);
Current, 4 digits (1pA max.).
Measurement Resolution :
Voltage, 4*1/2 digits (1 mV max.);
Current, 4 digits (50fA max.).
Ranging :
Automatic.
Current/Voltage Limiting (Compliance) :
Current output from an SMU
operating as a voltage source and
voltage output from an SMU
operating as a current source can be
limited.
Compliance Range : Current, 50pA to maximum allowable output current of
each voltage range; Voltage, OV to maximum allowable
output voltage of each current range.
Accuracy: Current compliance, accuracy of current source+1 % of range +lOpA;
Voltage compliance, accuracy of voltage source.
Residual Resistance (Voltage Source/Current Measurement Mode) : 0.452.
Input Resistance (Current Source/Voltage Measurement Mode) : Z- 101k.
Capacitive Load : 6 1 OOOpF.
l-6