HP 416A - Electrical Zero Adjustment

HP 416A
109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Q)
:>
0
,..0
ctj
7
'"d
~
ctj
a-
> I:'-
......
ctj
·
.-<
I-<
Q)
(/).
1.1")
1.1")
'-.......
N
'-.......
0
......
......
i
""'
0
0
<C
......
""'
3.
Connect
a
DC
vacuum
tube
voltmeter
,
such
as
a
Hewlett
-
Packard
Model
410B,
to
the
negative
terminal
of
capacitor
C70
.
Capacitor
C70
is
located
on
the
bottom
deck
of
the
Model
4 l
6A
and
is
i
dentified
in
Fig.
21.
The
negative
terminal
of
C70
is
just
opposite
the
center
tap
of
variable
resistors
R
118,
(Fig.
21).
Between
the
negative
terminal
of
C70
and
ground
should
meas
-
ure
-
6.
3V
with
a
line
voltage
of
l
15V
AC
.
If
this
indication
is
not
obtained
on
the
Model
410B,
variable
resistor
Rll8
may
be
adjusted
to
obtain
the
required
-
6.
3
volts.
Adjustment
of
HV
Output
Level
-
1.
Connect
a
DC
vacuum
tube
voltmeter
such
as
a
Hewlett
-
Packard
Model
410B
to
pin
3
of
Vl9
(Fig
.
21)
.
Between
pin
3
of
Vl9
and
ground
should
measure
+250
volts
with
a
line
voltage
of
l
15V
AC.
2.
If
this
indication
is
not
obtained
on
the
voltmeter,
variable
resistor
Rll6
(Fig.
21)
may
be
adjusted
to
obtain
the
required
+
250
volts
.
Line
Regulation
-
1.
Set
the
variable
transformer
so
that
the
line
voltage
to
the
Model
416A
will
be
at
103
.
5V
AC
and
then
at
126
.
5V
AC.
2.
The
indication
on
the
DC
vacuum
tube
voltmeter
connected
to
pin
3
of
Vl9
should
remain
at
+
250
volts
regardless
of
variations
between
103,
5
and
12
6.
5
volts
in
the
power
supply
input
.
3.
If
the
regulator
output
voltage
varies
with
input
voltage
variations,
see
Trouble
-
Shooting
Chart
,
step
2
under
"HV
Regulator
not
functioning".
Calibration
-
The
calibration
procedure
consists
of
adjusting
the
meter
circuit
for
electrical
zero
and
electrical
full
scale
.
Remove
the
cab
inet
from
the
Model
416A
before
starting
the
calibration
procedure
.
Test
Set
-
Up
: -
For
the
calibration
adjustments
,
equipment
shou
ld
be
arranged
as
shown
in
Fig
.
18
.
Note
that
the
cable
which
extends
the
connection
from
attenuator
#2
to
the
REFLECTED/PROBE
connector
on
the
Model
416A
is
fitted
with
a
shielded
dual
banana
plug
.
The
res
i
stor
across
the
output
of
attenuator
#2
may
be
connected
ins
i
de
the
connector
shield
as
shown
in
De
-
tail
1.
To
avoid
injection
of
noise
from
external
sources
into
the
low
-
noise
amplifier
circuits
of
the
Model
416A
,
the
units
shown
in
Fig
.
18
should
not
be
connected
to
an
existing
bench
or
test
stand
ground
.
Electrical
Zero
Adjustment
:
1.
The
input
to
attenuator
#1
is
to
be
10
volts
at
1
kc
.
Set
the
audio
oscillator
for
an
output
of
1
kc
at
such
level
that
a IO -
volt
reading
is
obtained
on
the
voltmeter
connected
across
the
input
to
attenuator
#1.
2.
At
the
Model
416A,
disconnect
the
reflected
input
cable
from
the
REFLECT
ED
/PROBE
connector
.
-
52
-

Related product manuals